超声杂波幅度、影响及空间范围的定量评估。

Quantitative assessment of the magnitude, impact and spatial extent of ultrasonic clutter.

作者信息

Lediju Muyinatu A, Pihl Michael J, Dahl Jeremy J, Trahey Gregg E

机构信息

Department of Biomedical Engineering, Duke University, Durham, NC, USA.

出版信息

Ultrason Imaging. 2008 Jul;30(3):151-68. doi: 10.1177/016173460803000302.

Abstract

Clutter is anoise artifact in ultrasound images that appears as diffuse echoes overlying signals of interest. It is most easily observed in anechoic or hypoechoic regions, such as in cysts, blood vessels, amniotic fluid, and urine-filled bladders. Clutter often obscures targets of interest and complicates anatomical measurements. An analytical expression that characterizes the extent to which clutter degrades lesion contrast was derived and compared to the measured contrast loss due to clutter in a bladder phantom. Simulation and phantom studies were performed to determine ideal and achievable signal-to-clutter ratios. In vivo clutter magnitudes were quantified in simultaneously-acquired fundamental and harmonic bladder images from five volunteers. Clutter magnitudes ranged from -30 dB to 0 dB, relative to the mean signal of the bladder wall. For this range of clutter magnitudes, the analytical expression predicts a contrast loss of 0-45 dB for lesions with clutter-free contrasts of 6-48 dB. A pixel-wise comparison of simultaneously-acquired fundamental and harmonic bladder images from each volunteer revealed an overall signal reduction in harmonic images, with average reductions ranging from 11-18 dB in the bladder interior and 9-11 dB in the tissue surrounding the bladder. Harmonic imaging did not reduce clutter in all volunteers.

摘要

杂波是超声图像中的一种噪声伪像,表现为覆盖在感兴趣信号上的弥漫性回声。在无回声或低回声区域最容易观察到,如囊肿、血管、羊水和充满尿液的膀胱中。杂波常常会掩盖感兴趣的目标,并使解剖测量变得复杂。推导了一个表征杂波降低病变对比度程度的解析表达式,并将其与膀胱模型中由于杂波导致的测量对比度损失进行了比较。进行了模拟和模型研究以确定理想的和可实现的信杂比。对五名志愿者同时采集的膀胱基波和谐波图像中的体内杂波幅度进行了量化。相对于膀胱壁的平均信号,杂波幅度范围为-30dB至0dB。对于这个杂波幅度范围,解析表达式预测,对于无杂波对比度为6-48dB的病变,对比度损失为0-45dB。对每位志愿者同时采集的膀胱基波和谐波图像进行逐像素比较,结果显示谐波图像中的信号总体减少,膀胱内部平均减少11-18dB,膀胱周围组织平均减少9-11dB。谐波成像并没有减少所有志愿者的杂波。

相似文献

2
A motion-based approach to abdominal clutter reduction.基于运动的腹部杂波减少方法。
IEEE Trans Ultrason Ferroelectr Freq Control. 2009 Nov;56(11):2437-49. doi: 10.1109/TUFFc.2009.1331.
3
Reverberation clutter from subcutaneous tissue layers: simulation and in vivo demonstrations.皮下组织层的混响杂波:模拟与体内演示
Ultrasound Med Biol. 2014 Apr;40(4):714-26. doi: 10.1016/j.ultrasmedbio.2013.11.029. Epub 2014 Feb 14.
5
Noninvasive bladder volume measurements based on nonlinear wave distortion.
Ultrasound Med Biol. 2004 Apr;30(4):469-76. doi: 10.1016/j.ultrasmedbio.2003.11.006.
7
Iterative Model-Based Beamforming for High Dynamic Range Applications.迭代模型基波束形成在高动态范围应用中的应用。
IEEE Trans Ultrason Ferroelectr Freq Control. 2021 Mar;68(3):482-493. doi: 10.1109/TUFFC.2020.3012165. Epub 2021 Feb 25.
8
9
Effect of dual-frequency clutter suppression in harmonic Doppler detection.双频杂波抑制对谐波多普勒检测的影响。
Ultrasonics. 2012 Feb;52(2):238-43. doi: 10.1016/j.ultras.2011.08.008. Epub 2011 Aug 25.

引用本文的文献

2
In Vivo Cavitation-Based Aberration Correction of Histotripsy in Porcine Liver.体内空化校正对猪肝脏组织微泡破碎的影响
IEEE Trans Ultrason Ferroelectr Freq Control. 2024 Aug;71(8):1019-1029. doi: 10.1109/TUFFC.2024.3409638. Epub 2024 Aug 19.
3
In Vivo Demonstration of a Real-Time Temporal SNR Acoustic Output Adjustment Method.体内实时时间信噪比声输出调整方法的验证。
IEEE Trans Ultrason Ferroelectr Freq Control. 2024 Aug;71(8):960-971. doi: 10.1109/TUFFC.2024.3402530. Epub 2024 Aug 19.
9
Combining ADMIRE and MV to Improve Image Quality.结合 ADMIRE 和 MV 以提高图像质量。
IEEE Trans Ultrason Ferroelectr Freq Control. 2022 Sep;69(9):2651-2662. doi: 10.1109/TUFFC.2022.3194548. Epub 2022 Aug 26.
10
Adaptive Models for Multi-Covariate Imaging of Sub-Resolution Targets (MIST).多变量亚分辨率目标成像(MIST)的自适应模型。
IEEE Trans Ultrason Ferroelectr Freq Control. 2022 Jul;69(7):2303-2317. doi: 10.1109/TUFFC.2022.3178035. Epub 2022 Jun 30.

本文引用的文献

5
Stationary clutter rejection in echocardiography.超声心动图中的固定杂波抑制
Ultrasound Med Biol. 2006 Jan;32(1):43-52. doi: 10.1016/j.ultrasmedbio.2005.08.012.
6
Improvement of tissue harmonic imaging using the pulse-inversion technique.使用脉冲反转技术改善组织谐波成像。
Ultrasound Med Biol. 2005 Jul;31(7):889-94. doi: 10.1016/j.ultrasmedbio.2005.03.006.
7
Performance testing of medical ultrasound equipment: fundamental vs. harmonic mode.
Ultrasonics. 2002 May;40(1-8):585-91. doi: 10.1016/s0041-624x(02)00177-4.
8
Real-time spatial compound imaging in breast ultrasound.乳腺超声中的实时空间复合成像
Ultrasound Med Biol. 2002 Feb;28(2):155-63. doi: 10.1016/s0301-5629(01)00490-2.
9
Clutter filter design for ultrasound color flow imaging.用于超声彩色血流成像的杂波滤波器设计
IEEE Trans Ultrason Ferroelectr Freq Control. 2002 Feb;49(2):204-16. doi: 10.1109/58.985705.
10
Nonlinear acoustics in diagnostic ultrasound.诊断超声中的非线性声学
Ultrasound Med Biol. 2002 Jan;28(1):1-18. doi: 10.1016/s0301-5629(01)00463-x.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验