Quality assurance of an image guided intracranial stereotactic positioning system.

作者信息

Soisson Emilie T, Sobering Geoff, Lucas Daniel, Chao Edward, Olivera Gustavo, Tomé Wolfgang A

机构信息

Department of Human Oncology, University of Wisconsin School of Medicine and Public Health, Madison, WI, USA.

出版信息

Technol Cancer Res Treat. 2009 Feb;8(1):39-49. doi: 10.1177/153303460900800106.

Abstract

This work reports on the development and testing of an intracranial stereotactic patient positioning system (ISPPS) for Tomotherapy. The ISPPS consists of the combination of a head frame, head frame couch interface (HCI), megavoltage CT (MVCT), and optical tracking camera system. Three quality assurance tests were designed to quantify the positioning system's ability to localize an intracranial target. The first two of these tests were designed to determine (a) the ability of the MVCT to detect a known shift applied to an anthropomorphic phantom and (b) the precision of fixing the phantom to the treatment couch via a head frame and specially designed head frame couch interface. A system verification test, using a phantom and EDR2 film, was used to determine the overall delivery precision through comparison of a measured dose distribution on film to calculated dose. The average net translational difference between a known shift applied to a phantom and that detected by MVCT image fusion was 0.62 mm. Setup reproducibility of the head frame was measured with both MVCT and optical tracking. The frame setup precision was found to be well within 1 mm for translations as well as rotations. A system delivery verification test in phantom using film showed spatial agreement between planned and delivered dose distributions to within 1 mm.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验