Shen Ya, Coil Jeffrey M, Haapasalo Markus
Department of Oral Biological and Medical Sciences, Division of Endodontics, University of British Columbia, Vancouver, Canada.
J Endod. 2009 Feb;35(2):193-6. doi: 10.1016/j.joen.2008.11.003. Epub 2008 Dec 13.
The purpose of this study was to analyze the incidence and mode of ProFile (Dentsply Tulsa Dental, Johnson City, TN) instrument separation during a predefined schedule of clinical use by the undergraduate students in a dental school over 4 years. A total of 3,706 ProFile instruments discarded from the same undergraduate students program between 2003 and 2007 were analyzed. The lateral and fracture surfaces of 12 separated instruments were examined by scanning electron microscopy, and the location of the fractures was recorded. The overall proportion of instrument defects was 1.3%; deformation without fracture occurred in 1% and separation in 0.3%. The majority of instrument defects occurred in size 20 (34/48). The ProFile instruments (10/12) failed mostly because of shear stress, whereas only two failed because of fatigue fracture. The results of this study indicated that NiTi rotary instrument system was successfully introduced into an undergraduate endodontic program. Small size files should be considered as single-use, disposable instruments because of the higher possibility of torsional deformation.
本研究的目的是分析在一所牙科学院的本科学生4年临床使用预定计划期间,Profile(登士柏塔尔萨牙科公司,田纳西州约翰逊城)器械分离的发生率和模式。对2003年至2007年间同一本科学生项目中丢弃的总共3706支Profile器械进行了分析。通过扫描电子显微镜检查了12支分离器械的侧面和断裂面,并记录了断裂位置。器械缺陷的总体比例为1.3%;1%发生了无断裂的变形,0.3%发生了分离。大多数器械缺陷发生在20号尺寸(48支中有34支)。Profile器械(12支中有10支)大多因剪切应力而失效,而只有两支因疲劳断裂而失效。本研究结果表明,镍钛旋转器械系统已成功引入本科牙髓病学项目。由于扭转变形的可能性较高,小尺寸锉应被视为一次性使用器械。