Kuster Markus S, Jeffcote Benjamin O, Schirm Andreas C, Jacob Hilaire, Nicholls Rochelle L
Fremantle Orthopaedic Unit, The University of Western Australia, Australia.
Knee. 2009 Oct;16(5):352-7. doi: 10.1016/j.knee.2009.01.002. Epub 2009 Feb 11.
Total knee arthroplasty (TKA) relies on soft tissue to regulate joint stability after surgery. In practice, the exact balance of the gaps can be difficult to measure, and various methods including intra-operative spreaders or distraction devices have been proposed. While individual ligament strain patterns have been measured, no data exist on the isometricity of the soft tissue envelope as a whole. In this study, a novel device was developed and validated to compare isometricity in the entire soft tissue envelope for both the intact and TKA knee. A spring-loaded rod was inserted in six cadaver knee joints between the tibial shaft and the tibial plateau or tibial tray after removing a 7 mm slice of bone. The displacement of the rod during passive flexion represented variation in tissue tension around the joint. The rod position in the intact knee remained within 1 mm of its initial position between 15 degrees and 135 degrees of flexion, and within 2 mm (+/-1.2 mm) throughout the entire range of motion (0-150 degrees). After insertion of a mobile-bearing TKA, the rod was displaced a mean of 6 mm at 150 degrees (p<0.001). The results were validated using a force transducer implanted in the tibial baseplate of the TKA, which showed increased tibiofemoral force in the parts of the flexion range where the rod was most displaced. The force measurements were highly correlated with the displacement pattern of the spring-loaded rod (r=-0.338; p=0.006). A simple device has been validated to measure isometricity in the soft tissue envelope around the knee joint. Isometricity measurements may be used in the future to improve implantation techniques during TKA surgery.
全膝关节置换术(TKA)术后依靠软组织来调节关节稳定性。在实际操作中,间隙的精确平衡可能难以测量,人们已经提出了包括术中撑开器或牵张装置在内的各种方法。虽然已经测量了各个韧带的应变模式,但关于整个软组织包膜的等长性尚无数据。在本研究中,开发并验证了一种新型装置,用于比较完整膝关节和TKA膝关节整个软组织包膜的等长性。在去除7毫米厚的骨片后,将一个弹簧加载杆插入六个尸体膝关节的胫骨干与胫骨平台或胫骨托之间。被动屈曲过程中杆的位移代表关节周围组织张力的变化。在完整膝关节中,杆的位置在15度至135度屈曲之间保持在其初始位置的1毫米以内,在整个运动范围(0 - 150度)内保持在2毫米(±1.2毫米)以内。插入活动轴承TKA后,杆在150度时平均位移6毫米(p<0.001)。使用植入TKA胫骨基板的力传感器对结果进行了验证,结果显示在杆位移最大的屈曲范围内,胫股力增加。力的测量结果与弹簧加载杆的位移模式高度相关(r = -0.338;p = 0.006)。一种简单的装置已被验证可用于测量膝关节周围软组织包膜的等长性。等长性测量未来可能用于改进TKA手术中的植入技术。