In many X-ray clinics, the traditional photographic film has been replaced by an imaging plate (IP). The IP is re-usable and the purpose of this study was to test if image deterioration occurred after successive uses of the IP. The emphasis is placed on the efficiency of image formation and on image uniformity. In a cross-sectional study, 21 clinically used IPs were exposed with a standardized phantom imaging protocol. These IPs were in clinical use between one month and two years and the IPs were exposed between 191 and 3787 times. After digitizing, the mean pixel value (MPV) in a predefined image area was determined. The relation between MPV and IP uses was assessed. In a second experiment, image uniformity of 30 other clinically used IPs was visually inspected for artifacts on a diagnostic monitor. These IPs were in clinical use between one week and two years and exposed between 76 and 5373 times. The first experiment showed that no significant deterioration of the MPV with increasing usage count of the IP was present (p = 0.15). The second experiment showed the appearance of clinically relevant artifacts on the IP before 3000 uses. It was concluded that the efficiency of the image formation process does not significantly deteriorate after successive use of IPs and is therefore not expected to limit their life span. Mechanical handling in the digitizer of the used system seems to set a limit to IP durability. Uniformity should therefore be checked regularly in clinical quality control.
在许多 X 射线诊所中,传统的摄影胶片已经被成像板(IP)所取代。IP 是可重复使用的,本研究的目的是测试 IP 连续使用后是否会发生图像劣化。重点放在图像形成的效率和图像的均匀性上。在一项横断面研究中,使用标准化的体模成像方案对 21 个临床使用的 IP 进行了曝光。这些 IP 在临床使用了 1 个月至 2 年之间,使用次数在 191 到 3787 次之间。数字化后,确定了预定义图像区域中的平均像素值(MPV)。评估了 MPV 与 IP 使用次数之间的关系。在第二个实验中,使用诊断监视器对 30 个其他临床使用的 IP 的图像均匀性进行了目视检查,以查找伪影。这些 IP 在临床使用了 1 周至 2 年之间,使用次数在 76 到 5373 次之间。第一个实验表明,随着 IP 使用次数的增加,MPV 没有明显的恶化(p = 0.15)。第二个实验表明,在使用 3000 次之前,IP 上出现了临床相关的伪影。结论是,IP 连续使用后,图像形成过程的效率不会显著恶化,因此预计不会限制其使用寿命。使用系统的数字化仪中的机械处理似乎对 IP 的耐用性设置了限制。因此,应定期在临床质量控制中检查均匀性。