Berg Wolfgang, Bechler Robin, Laube Norbert
Department of Urology, Friedrich-Schiller-University of Jena, Jena, Germany.
Clin Chem Lab Med. 2009;47(4):478-82. doi: 10.1515/CCLM.2009.099.
Since its first publication in 2000, the BONN-Risk-Index (BRI) has been successfully used to determine the calcium oxalate (CaOx) crystallization risk from urine samples. To date, a BRI-measuring device, the "Urolizer", has been developed, operating automatically and requiring only a minimum of preparation. Two major objectives were pursued: determination of Urolizer precision, and determination of the influence of 24-h urine storage at moderate temperatures on BRI.
24-h urine samples from 52 CaOx stone-formers were collected. A total of 37 urine samples were used for the investigation of Urolizer precision by performing six independent BRI determinations in series. In total, 30 samples were taken for additional investigation of urine storability. Each sample was measured thrice: directly after collection, after 24-h storage at T=21 degrees C, and after 24-h cooling at T=4 degrees C. Outcomes were statistically tested for identity with regard to the immediately obtained results.
Repeat measurements for evaluation of Urolizer precision revealed statistical identity of data (p-0.05). 24-h storage of urine at both tested temperatures did not significantly affect BRI (p-0.05).
The pilot-run Urolizer shows high analytical reliability. The innovative analysis device may be especially suited for urologists specializing in urolithiasis treatment. The possibility for urine storage at moderate temperatures without loss of analysis quality further demonstrates the applicability of the BRI method.
自2000年首次发表以来,波恩风险指数(BRI)已成功用于从尿液样本中确定草酸钙(CaOx)结晶风险。迄今为止,已开发出一种BRI测量设备“Urolizer”,其能自动运行且仅需最少的准备工作。本研究有两个主要目标:确定Urolizer的精度,以及确定在中等温度下储存24小时尿液对BRI的影响。
收集了52名草酸钙结石形成者的24小时尿液样本。总共37份尿液样本用于通过连续进行六次独立的BRI测定来研究Urolizer的精度。总共采集了30份样本用于尿液可储存性的额外研究。每个样本测量三次:采集后立即测量、在21摄氏度下储存24小时后测量以及在4摄氏度下冷却24小时后测量。对结果进行统计学检验,以确定与立即获得的结果是否一致。
评估Urolizer精度的重复测量显示数据具有统计学一致性(p>0.05)。在两个测试温度下储存24小时的尿液对BRI没有显著影响(p>0.05)。
试运行的Urolizer显示出高分析可靠性。这种创新的分析设备可能特别适合专门从事尿石症治疗的泌尿科医生。在中等温度下储存尿液而不损失分析质量的可能性进一步证明了BRI方法的适用性。