Ebert M A, Howlett S J, Harrison K, Cornes D, Hamilton C S, Denham J D
Department of Radiation Oncology, Sir Charles Gairdner Hospital, Nedlands, Australia.
Australas Phys Eng Sci Med. 2008 Dec;31(4):268-79. doi: 10.1007/BF03178596.
This paper describes the process and results of a survey of linear accelerator outputs as part of an Australasian Level III Dosimetry Intercomparison. This study involved the measurement of accelerator output under reference conditions ('Level I') with a small-volume ionisation chamber in water for 47 beams at 36 radiotherapy centres using the IAEA TRS 398 dose-to-water protocol. The mean ratio of measured to locally-determined accelerator output was 1.003 +/- 0.009 (1 standard deviation) with a range from 0.981 to 1.024. No correlation could be found between output ratio and accelerator type or local output calibration protocol. The small-volume chamber used satisfied most requirements for the study though showed some variation in sensitivity via repeated cross-calibration with a chamber calibrated at a primary standards laboratory.
本文描述了作为澳大拉西亚三级剂量学比对一部分的直线加速器输出量调查的过程和结果。本研究涉及在参考条件(“一级”)下,使用小体积电离室在水中对36个放射治疗中心的47束射线进行加速器输出量测量,采用国际原子能机构TRS 398水吸收剂量规程。测量值与本地确定的加速器输出量的平均比值为1.003±0.009(1个标准差),范围为0.981至1.024。未发现输出量比值与加速器类型或本地输出量校准规程之间存在相关性。所使用的小体积电离室满足了本研究的大部分要求,不过通过与在一级标准实验室校准的电离室进行重复交叉校准,其灵敏度显示出一定变化。