Oliva Alexis, Llabrés Matías, Fariña José B
Dpto. Ingeniería Química y Tecnología Farmacéutica, Facultad de Farmacia, Universidad de La Laguna, Avda. Fco. Sanchez, s/n, La Laguna, 38200, La Laguna, Tenerife, Spain.
Talanta. 2009 May 15;78(3):781-9. doi: 10.1016/j.talanta.2008.12.039. Epub 2008 Dec 30.
Size-exclusion chromatography (SEC) coupled with online laser light-scattering (LS) and refractive index (RI) detection provides an excellent approach to determine the molecular weights (Mw) of proteins by the "two-detector" approach. Mw is determined only at the maximum of a peak, using either peak heights or area ratio from the two detectors. However, proper calibration of the SEC/LS/RI system is critical to obtain high precision. Today, an essential part of any analysis is to evaluate the uncertainty associated with the method. Basically, it is possible to distinguish between factors related to signal nature, precision and those due to signal processing. Given the signal of interest is the peak height or area ratio from two detectors, the signal ratio uncertainty was calculated using the random propagation of error formula. In this case, the effect of signal correlation was evaluated to avoid the uncertainty overestimation. In the second case, the sources of uncertainty affecting analytical measurement were estimated with the information from the precision assessment. For this, two designs with two-factor fully nested were followed for each method. Finally, the contributions from various uncertainty sources related with calibration are also analysed in detail. There are in fact only three main sources of measurement uncertainty: intermediate precision, calibration and repeatability. Of these, method precision is always the greatest, regardless of approach. For all proteins and peptides studied, the Mw calculated using both methods are close to the theoretical results, independently of the design, but the contributions of individual terms to combined uncertainty depend on both the design and method used. For example, the combined uncertainty varied between 223 and 813.2 Da for carbonic anhydrase, although higher values were found for human insulin and ovalbumin dimer. Other considerations that can have a significant impact on the results are discussed. The reproducibility of the two methods versus that based on ASTRA software used as reference method was performed using the concordance correlation coefficient. The methods' reproducibility depends on the permitted losses in precision and accuracy.
尺寸排阻色谱法(SEC)与在线激光散射(LS)和折射率(RI)检测相结合,通过“双检测器”方法为测定蛋白质的分子量(Mw)提供了一种出色的途径。Mw仅在峰的最大值处测定,使用两个检测器的峰高或面积比。然而,SEC/LS/RI系统的正确校准对于获得高精度至关重要。如今,任何分析的一个重要部分是评估与该方法相关的不确定度。基本上,可以区分与信号性质、精密度相关的因素以及与信号处理相关的因素。鉴于感兴趣的信号是两个检测器的峰高或面积比,使用误差公式的随机传播来计算信号比不确定度。在这种情况下,评估信号相关性的影响以避免不确定度高估。在第二种情况下,利用精密度评估的信息估计影响分析测量的不确定度来源。为此,每种方法都采用了双因素完全嵌套的两种设计。最后,还详细分析了与校准相关的各种不确定度来源的贡献。实际上,测量不确定度只有三个主要来源:中间精密度、校准和重复性。在这些来源中,无论采用何种方法,方法精密度始终是最大的。对于所有研究的蛋白质和肽,使用两种方法计算得到的Mw均接近理论结果,与设计无关,但各个项对合成不确定度的贡献取决于所使用的设计和方法。例如,碳酸酐酶的合成不确定度在223至813.2 Da之间变化,尽管人胰岛素和卵清蛋白二聚体的值更高。还讨论了其他可能对结果产生重大影响的因素。使用一致性相关系数对这两种方法与基于ASTRA软件用作参考方法的重现性进行了比较。方法的重现性取决于允许的精密度和准确度损失。