采用带电晕荷电气溶胶检测器的尺寸排阻色谱法分析聚乙二醇聚合物。

Size exclusion chromatography with Corona charged aerosol detector for the analysis of polyethylene glycol polymer.

作者信息

Kou Dawen, Manius Gerald, Zhan Shangdong, Chokshi Hitesh P

机构信息

Department of Pharmaceutical and Analytical Research & Development, Hoffmann-La Roche Inc., 340 Kingsland Street, Nutley, NJ 07110, USA.

出版信息

J Chromatogr A. 2009 Jul 10;1216(28):5424-8. doi: 10.1016/j.chroma.2009.05.043. Epub 2009 May 21.

Abstract

A technique of using size exclusion chromatography (SEC) with the Corona charged aerosol detector (CAD) was developed and evaluated in comparison with refractive index (RI) and evaporative light scattering detection (ELSD) for fast screening of polyethylene glycol (PEG), a polymer used in preparing pegylated pharmaceutical compounds. These detection techniques were used in the analysis of multiple lots of PEG reagents. CAD was found to provide more accurate impurity and polydispersity profiles of PEG reagents that better differentiate their quality, while RI was not suitable for this application due to its low sensitivity and ELSD led to underestimation of the impurity and polydispersity. The accuracy of polydispersity determination by SEC-CAD was validated against a commercial reference standard of known polydispersity. The SEC-CAD technique and the observed differences between the three detectors can also be applied to polymer analysis in general.

摘要

开发了一种使用尺寸排阻色谱法(SEC)与电晕带电气溶胶检测器(CAD)相结合的技术,并与折光率(RI)和蒸发光散射检测(ELSD)进行了比较评估,用于快速筛选聚乙二醇(PEG),这是一种用于制备聚乙二醇化药物化合物的聚合物。这些检测技术用于分析多个批次的PEG试剂。结果发现,CAD能够提供更准确的PEG试剂杂质和多分散性图谱,能更好地区分其质量,而RI由于灵敏度低不适合此应用,ELSD则导致杂质和多分散性被低估。通过SEC-CAD测定多分散性的准确性已根据已知多分散性的商业参考标准进行了验证。SEC-CAD技术以及观察到的三种检测器之间的差异通常也可应用于聚合物分析。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验