Lu Yanyan, Wang Wenjun, Xing Zhi, Wang Shidong, Cao Po, Zhang Sichun, Zhang Xinrong
Department of Chemistry, Key Laboratory for Atomic and Molecular Nanosciences of the Education Ministry, Tsinghua University, Beijing 100084, PR China.
Talanta. 2009 May 15;78(3):869-73. doi: 10.1016/j.talanta.2008.12.065. Epub 2009 Jan 20.
A sandwich-type immunoassay linked with inductively coupled plasma mass spectrometry (ICP-MS) has been developed for the detection of anti-erythropoietin antibodies (anti-EPO Abs). Recombinant human erythropoietin (rhEPO) was immobilized on the solid phase to capture anti-rhEPO Abs specifically. After the immunoreactions with Au-labeled goat-anti-rabbit IgG, a diluted HNO(3) (2%) was used to dissociate Au nanoparticles which was then introduced to the ICP-MS for measurements. Under the optimized conditions, the calibration graph for anti-EPO Abs was linear in the range of 35.6-500 ng mL(-1) with a detection limit of 10.7 ng mL(-1) (3sigma, n=9). The relative standard deviation (R.S.D.) for three replicate measurements of 30.9 ng mL(-1) of anti-EPO Abs was 8.43%. The recoveries of anti-EPO Abs in sera at the spiking level of 50, 100, 150, 200 and 400 ng mL(-1) were 99.2%, 101.5%, 95.0%, 94.0% and 102.9%, respectively. For the real sample analysis, 26 samples from healthy people and 53 samples from patients with rhEPO treatments were studied. One sample from patients showed significantly higher anti-EPO Abs from other samples, indicating a possibility of immune response of this patient.
一种与电感耦合等离子体质谱法(ICP-MS)联用的夹心型免疫分析法已被开发用于检测抗促红细胞生成素抗体(抗EPO抗体)。重组人促红细胞生成素(rhEPO)被固定在固相上以特异性捕获抗rhEPO抗体。在与金标记的山羊抗兔IgG进行免疫反应后,使用稀释的HNO₃(2%)解离金纳米颗粒,然后将其引入ICP-MS进行测量。在优化条件下,抗EPO抗体的校准曲线在35.6 - 500 ng mL⁻¹范围内呈线性,检测限为10.7 ng mL⁻¹(3σ,n = 9)。对30.9 ng mL⁻¹的抗EPO抗体进行三次重复测量的相对标准偏差(R.S.D.)为8.43%。在加标水平为50、100、150、200和400 ng mL⁻¹时,血清中抗EPO抗体的回收率分别为99.2%、101.5%、95.0%、94.0%和102.9%。对于实际样品分析,研究了26份来自健康人的样品和53份接受rhEPO治疗患者的样品。一名患者的样品显示其抗EPO抗体明显高于其他样品,表明该患者可能存在免疫反应。