Narracott A J, John G W, Morris R J, Woodcock J P, Hose D R, Lawford P V
Academic Unit of Medical Physics, School of Medicine and Biomedical Sciences, University of Sheffield, Sheffield S10 2RX, UK.
IEEE Trans Biomed Eng. 2009 Feb;56(2):273-80. doi: 10.1109/TBME.2008.2005964. Epub 2008 Oct 3.
This paper presents a validated model of calf compression with an external pressure cuff as used for deep vein thrombosis. Magnetic resonance (MR) images of calf geometry were used to generate subject-specific finite-element (FE) models of the calf cross section. Ultrasound images of deep vessel collapse obtained through a water-filled cuff were used to validate model behavior. Calf/cuff pressure interface measurements were applied to the FE model and the resulting tissue deformation was compared with MR image in normal volunteers (three females, four males, age range 20-55) using two distinct cuffs. MR observations and the model results showed good qualitative agreement. A similar reduction in cross-sectional area of the posterior tibial veins was obtained under both symmetric compression (89%) and asymmetric compression (81%), but greater compression of the anterior tibial veins was achieved with symmetric compression. The need to account for the effective compressibility of the calf tissue suggests that external measurements of the calf tissue deformation will not accurately predict deep vessel collapse. These results have implications for the modification of venous haemodynamics by such systems and could help to improve cuff design.
本文介绍了一种经过验证的用于深静脉血栓形成的外部压力袖带小腿压迫模型。利用小腿几何结构的磁共振(MR)图像生成小腿横截面的个体有限元(FE)模型。通过充水袖带获得的深部血管塌陷超声图像用于验证模型行为。将小腿/袖带压力界面测量应用于有限元模型,并使用两种不同的袖带,将正常志愿者(三名女性,四名男性,年龄范围20 - 55岁)的组织变形结果与MR图像进行比较。MR观察结果与模型结果显示出良好的定性一致性。在对称压迫(89%)和不对称压迫(81%)下,胫后静脉的横截面积都有类似程度的减小,但对称压迫对胫前静脉的压迫更大。考虑小腿组织有效可压缩性的必要性表明,小腿组织变形的外部测量不能准确预测深部血管塌陷。这些结果对这类系统改变静脉血流动力学有影响,并有助于改进袖带设计。