Kurt Rossmann Laboratories for Radiologic Image Research, MC2026, Department of Radiology, The University of Chicago, 5841 S. Maryland Avenue, Chicago, IL 60637, USA.
J Digit Imaging. 2010 Apr;23(2):161-9. doi: 10.1007/s10278-009-9192-x. Epub 2009 Mar 10.
We evaluated the potential utility of a newly developed liquid-crystal display (LCD), which used an independent sub-pixel drive (ISD) technique for increasing the spatial resolution of a standard LCD three times in one direction, by use of receiver operating characteristic (ROC) analysis and a two-alternative-forced-choice (2AFC) method to determine improvement in radiologists' accuracy in the detection of clustered microcalcifications (MCLs) on digital mammograms. We used a standard LCD without and with the ISD technique, which can increase the spatial resolution of the LCD three times in one direction from three mega- to nine megapixels without changes in the size of the display. We used 60 single views of digital mammograms (30 with and 30 without clustered MCLs) for ROC studies and 60 regions of interest (ROIs) with clustered MCLs for 2AFC studies. In the ROC study, seven radiologists attempted to detect clustered MCLs without and with the ISD on the same LCD. In the 2AFC study, the same observer group compared the visibility of MCLs by use of the LCD without and with the ISD. Our institutional review board approved the use of this database and the participation of radiologists in this study. The accuracy in detecting clustered MCLs in the ROC study was improved by use of the LCD with the ISD, but the improvement was not statistically significant (p = 0.08). However, the superiority of the LCD with the ISD was demonstrated as significant (p < 0.001) in the 2AFC study. An LCD with ISD can improve the visibility of clustered MCLs when high-resolution digital mammograms are available.
我们评估了一种新开发的液晶显示器(LCD)的潜在效用,该显示器使用独立子像素驱动(ISD)技术将标准 LCD 的空间分辨率在一个方向上提高三倍,通过使用接收器操作特性(ROC)分析和二选一强制选择(2AFC)方法来确定放射科医生在检测数字乳房 X 线照片上簇状微钙化(MCL)的准确性是否提高。我们使用了标准的 LCD 和具有 ISD 技术的 LCD,该技术可以在不改变显示器尺寸的情况下将 LCD 的空间分辨率在一个方向上从三个兆像素提高到九个兆像素。我们使用了 60 张数字乳房 X 线照片的单视图(30 张有簇状 MCL 和 30 张没有簇状 MCL)进行 ROC 研究,以及 60 个具有簇状 MCL 的感兴趣区域(ROI)进行 2AFC 研究。在 ROC 研究中,七名放射科医生试图在同一台 LCD 上不使用和使用 ISD 来检测簇状 MCL。在 2AFC 研究中,同一观察组使用无 ISD 和有 ISD 的 LCD 来比较 MCL 的可见度。我们的机构审查委员会批准了该数据库的使用和放射科医生参与这项研究。在 ROC 研究中,使用具有 ISD 的 LCD 可以提高检测簇状 MCL 的准确性,但这种提高没有统计学意义(p = 0.08)。然而,在 2AFC 研究中,ISD 显示器的优越性被证明是显著的(p < 0.001)。当有高分辨率数字乳房 X 线照片时,具有 ISD 的 LCD 可以提高簇状 MCL 的可见度。