Effect of hydrodynamic parameters on corneal endothelial cell loss after phacoemulsification.

作者信息

Baradaran-Rafii Alireza, Rahmati-Kamel Mohsen, Eslani Medi, Kiavash Victoria, Karimian Farid

机构信息

Department of Ophthalmology and Ophthalmic Research Center, Labbafinejad Medical Center, Shahid Beheshti University, Tehran, Iran.

出版信息

J Cataract Refract Surg. 2009 Apr;35(4):732-7. doi: 10.1016/j.jcrs.2008.12.017.

Abstract

PURPOSE

To evaluate the effect of power, vacuum, and flow rate on endothelial cell loss after phacoemulsification.

SETTING

Labbafinejad Medical Center Medical Center, Tehran, Iran.

METHODS

In a prospective randomized clinical trial, phacoemulsification was performed in 2 groups (high vacuum and low vacuum) with 3+ nuclear sclerosis. The stop-and-chop technique was used with the Sovereign machine. Machine parameters during the chop stage were vacuum 400 mm Hg in the high-vacuum group and 200 mm Hg in the low-vacuum group and flow rate, 40 cc/min and 20 cc/min, respectively. Endothelial cell density preoperatively before and 1, 6, and 12 weeks postoperatively were compared.

RESULTS

Each group comprised 30 eyes. The mean US power was 9.2% +/- 4.3% (SD) in the low-vacuum group and 13.1% +/- 4.6% in the high-vacuum group (P = .001) and the mean phaco time, 1.28 +/- 1.0 minutes and 0.88 +/- 0.6 minutes, respectively (P = .04). Total US energy and total fluid consumed were similar between groups. After 12 weeks, the mean endothelial cell loss was 9.0% +/- 4.0% in the low-vacuum group and 9.6% +/- 4.6% in the high-vacuum group (P = .6). There was a relationship between total US energy and endothelial loss (P<.001); however, total fluid volume was not a significant predictor (P = .19).

CONCLUSIONS

Vacuum level did not have a significant effect on total US energy or total fluid consumed during phacoemulsification. There was a strong relationship between total US energy and endothelial cell loss but not between total infused fluid and endothelial cell loss.

摘要

目的

评估能量、真空度和流速对超声乳化术后内皮细胞损失的影响。

设置

伊朗德黑兰拉巴菲内贾德医疗中心。

方法

在一项前瞻性随机临床试验中,对两组(高真空组和低真空组)核硬度为3级的患者进行超声乳化手术。使用Sovereign机器采用拦截劈核技术。劈核阶段的机器参数为,高真空组真空度400毫米汞柱,低真空组200毫米汞柱,流速分别为40毫升/分钟和20毫升/分钟。比较术前及术后1周、6周和12周的内皮细胞密度。

结果

每组30只眼。低真空组平均超声能量为9.2%±4.3%(标准差),高真空组为13.1%±4.6%(P = 0.001),平均超声乳化时间分别为1.28±1.0分钟和0.88±0.6分钟(P = 0.04)。两组之间的总超声能量和总液体消耗量相似。12周后,低真空组平均内皮细胞损失为9.0%±4.0%,高真空组为9.6%±4.6%(P = 0.6)。总超声能量与内皮细胞损失之间存在相关性(P<0.001);然而,总液体量不是一个显著的预测因素(P = 0.19)。

结论

真空度对超声乳化过程中的总超声能量或总液体消耗量没有显著影响。总超声能量与内皮细胞损失之间存在密切关系,但总灌注液量与内皮细胞损失之间不存在这种关系。