Stepp Cara E, Heaton James T, Rolland Rebecca G, Hillman Robert E
Harvard-MIT Division of Health Sciences and Technology, Cambridge, MA 02139, USA.
IEEE Trans Neural Syst Rehabil Eng. 2009 Apr;17(2):146-55. doi: 10.1109/TNSRE.2009.2017805. Epub 2009 Mar 16.
The electrolarynx (EL) is a common rehabilitative speech aid for individuals who have undergone total laryngectomy, but they typically lack pitch control and require the exclusive use of one hand. The viability of using neck and face surface electromyography (sEMG) to control the onset, offset, and pitch of an EMG-controlled EL (EMG-EL) was studied. Eight individuals who had undergone total laryngectomy produced serial and running speech using a typical handheld EL and the EMG-EL while attending to real-time visual sEMG biofeedback. Running speech tokens produced with the EMG-EL were examined for naturalness by 10 listeners relative to those produced with a typical EL using a visual analog scale. Serial speech performance was assessed as the percentage of words that were fully voiced and pauses that were successfully produced. Results of the visual analog scale assessment indicated that individuals were able to use the EMG-EL without training to produce running speech perceived as natural as that produced with a typical handheld EL. All participants were able to produce running and serial speech with the EMG-EL controlled by sEMG from multiple recording locations, with the superior ventral neck or submental surface locations providing at least one of the two best control locations.
电子喉(EL)是一种常见的用于全喉切除患者的康复性言语辅助装置,但它通常缺乏音高控制功能,且需要单手操作。本研究探讨了利用颈部和面部表面肌电图(sEMG)来控制肌电图控制电子喉(EMG-EL)的启动、停止和音高的可行性。八名全喉切除患者在关注实时视觉sEMG生物反馈的同时,使用典型的手持式电子喉和EMG-EL进行连续和流畅言语表达。10名听众使用视觉模拟量表,将EMG-EL产生的流畅言语样本与典型电子喉产生的样本进行自然度比较。连续言语表现通过完全发声的单词百分比和成功产生的停顿来评估。视觉模拟量表评估结果表明,受试者无需训练就能使用EMG-EL进行流畅言语表达,且其流畅度与典型手持式电子喉相当。所有参与者都能够通过来自多个记录位置的sEMG控制EMG-EL进行流畅和连续言语表达,其中颈前腹侧或颏下表面位置至少是两个最佳控制位置之一。