Fresh gas utilization of eight circle systems.

作者信息

Zbinden A M, Feigenwinter P, Hutmacher M

机构信息

Institute for Anaesthesiology and Intensive Care, University Hospital, Bern, Switzerland.

出版信息

Br J Anaesth. 1991 Oct;67(4):492-9. doi: 10.1093/bja/67.4.492.

Abstract

The fresh gas utilization (FGU) of a semi-closed breathing system is defined as the ratio of the amount of gas reaching the patient's lungs to the total amount of fresh gas flowing into the breathing system. It indicates to what extent a breathing system conserves anaesthetic gases and provides inspired gas concentrations as close as possible to those in the fresh gas, even at low fresh gas flows (FGF). We have measured FGU in eight circle systems used conventionally in Europe: Dräger Cicero, Dräger Sulla 808V with circle system 8 ISO and ventilator Ventilog, Dräger AV1, Ohmeda Modulus II Plus, Gambro Engström Elsa, Siemens Servo Ventilator 900 D with circle system 985, Siemens Ventilator 710 and Megamed 700A with circle system 219. The Tests were performed according to the Draft European Standard "Anaesthetic Workstations and Their Modules". None of the systems tested showed the characteristics of an ideal system which would reach 100% FGU with an FGF less than minute volume. At FGF 3 litre min-1, FGU was: Gambro Engström Elsa 97.8%, Siemens Servo Ventilator 900 D with circle system 96.1%, Dräger Cicero 93.4%, Ohmeda Modulus II Plus 93.1%, Dräger 8 ISO 92.3%, Dräger AV1 87.6%, Megamed 700A 77.0% and Siemens Ventilator 710 74.1%.