Bhargava A K, Mahajan H K, Gandhi B K, Sharma S
Classified Specialist (Anaesthesiology), Department of Anaesthesiology, Command Hospital (NC), C/o 56 APO.
Graded Specialist (Anaesthesiology), Department of Anaesthesiology, Command Hospital (NC), C/o 56 APO.
Med J Armed Forces India. 1997 Oct;53(4):259-262. doi: 10.1016/S0377-1237(17)30749-9. Epub 2017 Jun 26.
The performance of a new anaesthesia machine was evaluated under low flow anaesthetic technique for 7 months. An assessment under low flow conditions over a definitive period can highlight the drawbacks or advantages of a machine under stringent conditions. The overall performance was satisfactory in the sense that it proved to be an efficient, reliable and safe machine. Minimum flows of oxygen and nitrous oxide that could be achieved at steady state were 0.4 L/min and 0.5 L/min respectively. At these flows the machine monitors were accurate and one could safely administer anaesthesia with a significant amount of saving of nitrous oxide. The requirement of nitrous oxide fell from 2 cylinders in a day to 2 cylinders in a week.
一台新型麻醉机在低流量麻醉技术下进行了7个月的性能评估。在一段确定的时间内进行低流量条件下的评估,可以凸显一台机器在严格条件下的缺点或优点。从总体性能令人满意这一点来说,它被证明是一台高效、可靠且安全的机器。稳态时可实现的氧气和一氧化二氮的最低流量分别为0.4升/分钟和0.5升/分钟。在这些流量下,机器监测准确,人们可以安全地实施麻醉,同时大量节省一氧化二氮。一氧化二氮的需求量从一天2瓶降至一周2瓶。