Suppr超能文献

低轮廓猪源Liotta生物瓣在二尖瓣位置的意外长期耐用性。

Unexpected long durability of a low-profile porcine Liotta bioprosthesis in mitral position.

作者信息

Campanella Antonio, Comoglio Chiara, Sansone Fabrizio, Rinaldi Mauro

机构信息

Division of Cardiovascular Surgery, University Hospital 'S. Giovanni Battista', Turin, Italy.

出版信息

J Cardiovasc Med (Hagerstown). 2009 Jun;10(6):492-3. doi: 10.2459/JCM.0b013e32832720a0.

Abstract

We report a case in which replacement of a low-profile porcine Liotta bioprosthesis was required 21 years after initial implantation. The patient underwent mitral valve replacement with a 28-mm Liotta bioprosthesis in 1985 for mitral stenosis. Twenty-one years later, severe mitral regurgitation was detected at echocardiography and mitral valve reoperation combined with tricuspid annuloplasty was successfully performed. Focal perforation of two cusps with consequent intravalvular leak on bioprosthesis was found. This is the longest durability report for a Liotta bioprosthesis in mitral position.

摘要

我们报告一例患者,在最初植入低轮廓猪源Liotta生物瓣膜21年后需要进行瓣膜置换。该患者于1985年因二尖瓣狭窄接受了28毫米Liotta生物瓣膜二尖瓣置换术。21年后,超声心动图检测到严重二尖瓣反流,成功进行了二尖瓣再次手术并联合三尖瓣环成形术。发现生物瓣膜两个瓣叶有局灶性穿孔并伴有瓣内漏。这是关于二尖瓣位Liotta生物瓣膜最长耐久性的报告。