Ellis Scott J, Hillstrom Howard, Cheng Richard, Lipman Joe, Garrison Glen, Deland Jonathan T
Hospital for Special Surgery, 535 East 70th Street, New York, NY, 10021, USA.
Foot Ankle Int. 2009 Apr;30(4):333-40. doi: 10.3113/FAI.2009.0333.
The goal of this study was to develop an accurate and reliable supine plantar pressure measurement apparatus that could potentially be used in the operating room to guide operative decision making. It was hypothesized that plantar pressures would be similar in supine and vertical posture trials.
A supine plantar pressure system was developed using the Pliance 32 sensor array (Novel, Munich, Germany). Accuracy was tested with a standardized manometer and the Trublu calibration system (Novel). Next, bilateral feet of 10 healthy, asymptomatic patients were tested in the apparatus by two separate investigators. Intraclass correlation coefficients (ICC) were calculated to determine intrarater and interrater reliability for parameters of average mean pressure (AMP), peak pressure (PP) and percentage of total force (PF). Finally, plantar pressures were tested in vertical posture and compared to supine parameters.
The sensor was both linear and accurate. The ICC values demonstrated that the sensor was reliable for AMP (0.66 to 0.93), PP (0.75 to 0.94), and PF (0.69 to 0.97). In general, AMP, PP, and PF values were significantly different at most anatomical regions for supine and vertical posture trials (p < 0.0001).
This intraoperative pressure measurement system provided an accurate, linear, and reliable method to measure plantar pressure parameters in the supine subject. These measures were similar in magnitude to vertical posture, but were statistically different.
This intraoperative plantar pressure system could guide operative reconstruction of foot deformities whose outcome depends on the distribution of pressures across the plantar foot.
本研究的目的是开发一种准确可靠的仰卧位足底压力测量装置,该装置有可能用于手术室以指导手术决策。研究假设仰卧位和直立位试验中的足底压力相似。
使用Pliance 32传感器阵列(德国慕尼黑Novel公司)开发了一种仰卧位足底压力系统。用标准压力计和Trublu校准系统(Novel公司)测试准确性。接下来,由两名独立的研究人员在该装置中对10名健康无症状患者的双足进行测试。计算组内相关系数(ICC)以确定平均平均压力(AMP)、峰值压力(PP)和总力百分比(PF)参数的评分者内和评分者间可靠性。最后,在直立位测试足底压力并与仰卧位参数进行比较。
该传感器具有线性且准确。ICC值表明该传感器对于AMP(0.66至0.93)、PP(0.75至0.94)和PF(0.69至0.97)是可靠的。一般来说,仰卧位和直立位试验在大多数解剖区域的AMP、PP和PF值有显著差异(p < 0.0001)。
这种术中压力测量系统提供了一种准确、线性且可靠的方法来测量仰卧位受试者的足底压力参数。这些测量值在大小上与直立位相似,但在统计学上有差异。
这种术中足底压力系统可指导足部畸形的手术重建,其结果取决于足底压力的分布。