In an intensive care unit, alarms are used to call attention to a patient, to alert a change in the patient's physiology, or to warn of a failure in a medical device; however, up to 94% of the alarms are false. Our purpose in this study was to identify a means of reducing the number of false alarms.
An observer recorded time-stamped information of alarms and the presence of health care team members in the patient room; each alarm response was classified as effective (action taken within 5 min), ineffective (no response to the alarm), and ignored (alarm consciously ignored or actively silenced).
During the 200-h study period, 1271 separate entries by an individual to the room being observed were recorded, 1214 alarms occurred and 2344 tasks were performed. On average, alarms occurred 6.07 times per hour and were active for 3.28 min per hour; 23% were effective, 36% were ineffective, and 41% were ignored. The median alarm duration was 17 s. A 14-s delay before alarm presentation would remove 50% of the ignored and ineffective alarms, and a 19-s delay would remove 67%. Suctioning, washing, repositioning, and oral care caused 152 ignored or ineffective ventilator alarms.
Introducing a 19-s alarm delay and automatically detecting suctioning, repositioning, oral care, and washing could reduce the number of ineffective and ignored alarms from 934 to 274. More reliable alarms could elicit more timely response, reduce workload, reduce noise pollution, and potentially improve patient safety.
在重症监护病房中,警报用于引起对患者的关注,提醒患者生理状态发生变化,或警告医疗设备出现故障;然而,高达94%的警报是误报。我们开展这项研究的目的是确定一种减少误报数量的方法。
一名观察者记录了带有时间戳的警报信息以及医护团队成员在病房内的情况;每个警报响应被分类为有效(在5分钟内采取行动)、无效(对警报无响应)和被忽略(警报被有意识地忽略或主动静音)。
在200小时的研究期间,记录了个人进入被观察病房的1271次单独记录,发生了1214次警报,执行了2344项任务。平均而言,每小时发生6.07次警报,每次警报持续3.28分钟;23%的警报响应有效,36%无效,41%被忽略。警报的中位持续时间为17秒。警报发出前延迟14秒将消除50%被忽略和无效的警报,延迟19秒将消除67%。吸痰、清洗身体、重新摆放体位和口腔护理导致152次被忽略或无效的呼吸机警报。
引入19秒的警报延迟并自动检测吸痰、重新摆放体位、口腔护理和清洗身体的操作,可以将无效和被忽略的警报数量从934次减少到274次。更可靠的警报可以引发更及时的响应,减轻工作量,减少噪音污染,并有可能提高患者安全。