Suppr超能文献

用于青光眼检测的图形视网膜电图(PERGLA)优化中操作者和光学离焦对变异性的影响。

Effect of operator and optical defocus on the variability of pattern electroretinogram optimized for glaucoma detection (PERGLA).

机构信息

Department of Ophthalmology, Hamilton Glaucoma Center, University of California, San Diego, La Jolla, CA 92037-0946, USA.

出版信息

J Glaucoma. 2010 Feb;19(2):77-82. doi: 10.1097/IJG.0b013e31819f934e.

Abstract

PURPOSE

To evaluate the effect of operator and optical defocus on the variability of pattern electroretinogram optimized for glaucoma detection (PERGLA).

METHODS

Two different operators obtained 2 PERGLA recordings each from 10 healthy participants (5 women, mean age 32.1+/-10.3 y). In addition, one of the operators obtained recordings in which corrective lenses of various diopters (+/-0.5, +/-1, +/-2, and +/-3) were used to generate optical defocus in both eyes. The effect of operator on PERGLA amplitude and phase variability was determined using a single nested variance components' analysis model and by using Bland-Altman plots. One-way analysis of variance (ANOVA) was used to determine the effect of optical defocus on amplitude and phase.

RESULTS

Differences in measurements between operators accounted for approximately 26.6% and 18.2% of the total variance for amplitude and phase, respectively. Results were confirmed by the use of Bland-Altman plots. ANOVA identified a significant effect of defocus on mean amplitude (F=2.65, P=0.01), but not phase (F=1.02, P=0.42).

CONCLUSIONS

Measurements obtained by different operators can result in significant differences in PERGLA amplitude. In addition, although optical defocus leads to a decrease in PERGLA amplitude by reducing visual acuity, this can be avoided by obtaining J1 or better vision before testing.

摘要

目的

评估操作者和光学离焦对青光眼检测优化图形视网膜电图(PERGLA)变异性的影响。

方法

两名不同的操作者分别对 10 名健康参与者(5 名女性,平均年龄 32.1+/-10.3 岁)进行了 2 次 PERGLA 记录。此外,其中一名操作者使用各种屈光度(+/-0.5、+/-1、+/-2 和+/-3)的矫正镜片在双眼产生光学离焦,从而获得记录。使用单嵌套方差分量分析模型和 Bland-Altman 图确定操作者对 PERGLA 幅度和相位变异性的影响。使用单因素方差分析(ANOVA)确定光学离焦对幅度和相位的影响。

结果

操作者之间的测量差异分别占幅度和相位总方差的约 26.6%和 18.2%。Bland-Altman 图也证实了这一结果。ANOVA 确定了离焦对平均幅度(F=2.65,P=0.01)有显著影响,但对相位(F=1.02,P=0.42)没有影响。

结论

不同操作者获得的测量结果可能导致 PERGLA 幅度的显著差异。此外,尽管光学离焦通过降低视力导致 PERGLA 幅度下降,但可以通过在测试前获得 J1 或更好的视力来避免这种情况。

相似文献

3
Reproducibility of pattern electroretinogram in glaucoma patients with a range of severity of disease with the new glaucoma paradigm.
Ophthalmology. 2008 Jun;115(6):957-63. doi: 10.1016/j.ophtha.2007.08.023. Epub 2007 Nov 5.
4
Diagnostic accuracy of pattern electroretinogram optimized for glaucoma detection.
Ophthalmology. 2009 Mar;116(3):437-43. doi: 10.1016/j.ophtha.2008.10.026. Epub 2009 Jan 22.
8
Pattern electroretinogram to detect glaucoma: comparing the PERGLA and the PERG Ratio protocols.
Doc Ophthalmol. 2013 Dec;127(3):227-38. doi: 10.1007/s10633-013-9412-z. Epub 2013 Oct 15.

引用本文的文献

1
Neuroprotection beyond intraocular pressure: game changer or quiet addiction.
Graefes Arch Clin Exp Ophthalmol. 2025 Apr 5. doi: 10.1007/s00417-025-06816-7.
2
Time-Frequency Analysis of ERG With Discrete Wavelet Transform and Matching Pursuits for Glaucoma.
Transl Vis Sci Technol. 2022 Oct 3;11(10):19. doi: 10.1167/tvst.11.10.19.
3
Can Variability of Pattern ERG Signal Help to Detect Retinal Ganglion Cells Dysfunction in Glaucomatous Eyes?
Biomed Res Int. 2015;2015:571314. doi: 10.1155/2015/571314. Epub 2015 Jun 8.
4
Pattern electroretinogram to detect glaucoma: comparing the PERGLA and the PERG Ratio protocols.
Doc Ophthalmol. 2013 Dec;127(3):227-38. doi: 10.1007/s10633-013-9412-z. Epub 2013 Oct 15.
5
Head-down posture induces PERG alterations in early glaucoma.
J Glaucoma. 2013 Mar;22(3):255-64. doi: 10.1097/IJG.0b013e318232973b.
6
Pattern electroretinogram association with spectral domain-OCT structural measurements in glaucoma.
Eye (Lond). 2011 Feb;25(2):224-32. doi: 10.1038/eye.2010.203. Epub 2010 Dec 24.

本文引用的文献

2
Diagnostic accuracy of pattern electroretinogram optimized for glaucoma detection.
Ophthalmology. 2009 Mar;116(3):437-43. doi: 10.1016/j.ophtha.2008.10.026. Epub 2009 Jan 22.
3
Update on the pattern electroretinogram in glaucoma.
Optom Vis Sci. 2008 Jun;85(6):386-95. doi: 10.1097/OPX.0b013e318177ebf3.
4
An improved method to continuously monitor the electrode-skin impedance during bioelectric measurements.
Annu Int Conf IEEE Eng Med Biol Soc. 2007;2007:6295-8. doi: 10.1109/IEMBS.2007.4353794.
5
Reproducibility of pattern electroretinogram in glaucoma patients with a range of severity of disease with the new glaucoma paradigm.
Ophthalmology. 2008 Jun;115(6):957-63. doi: 10.1016/j.ophtha.2007.08.023. Epub 2007 Nov 5.
7
Pattern electroretinogram in glaucoma.
Curr Opin Ophthalmol. 2006 Apr;17(2):196-202. doi: 10.1097/01.icu.0000193082.44938.3c.
8
Habituation of retinal ganglion cell activity in response to steady state pattern visual stimuli in normal subjects.
Invest Ophthalmol Vis Sci. 2005 Apr;46(4):1296-302. doi: 10.1167/iovs.04-1242.
9
Pattern electroretinogram abnormality and glaucoma.
Ophthalmology. 2005 Jan;112(1):10-9. doi: 10.1016/j.ophtha.2004.07.018.
10
Different effect of dioptric defocus vs. light scatter on the pattern electroretinogram (PERG).
Doc Ophthalmol. 2004 Jan;108(1):99-106. doi: 10.1023/b:doop.0000018415.00285.56.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验