Elsharkawy H, Lewis B S, Steiger E, Farag E
Institute of Anesthesiology, Cleveland Clinic, Cleveland, OH 44195, USA.
Minerva Anestesiol. 2009 Jul-Aug;75(7-8):471-4. Epub 2008 Jan 24.
Arrhythmias are common in hospitalized patients and during surgery. We present a case of positional atrial arrhythmia related to a peripherally inserted central catheter (PICC). There are other documented case reports of ventricular tachycardia precipitated by body position changes with a PICC. The immediate correction of the arrhythmia with repositioning of the PICC strongly points to the PICC as the cause. This highlights the potential seriousness of cardiac arrhythmias precipitated by PICCs as well as the need for careful catheter placement and perioperative maintenance. Practitioners should consider PICC line tip position as a rare cause of positional atrial arrhythmias.
心律失常在住院患者和手术期间很常见。我们报告一例与经外周静脉穿刺中心静脉导管(PICC)相关的体位性房性心律失常病例。有其他文献记载的病例报告显示,PICC导致体位改变诱发室性心动过速。通过重新放置PICC立即纠正心律失常有力地表明PICC是病因。这突出了PICC诱发心律失常的潜在严重性,以及导管放置和围手术期维护需谨慎的必要性。从业者应将PICC导管尖端位置视为体位性房性心律失常的罕见病因。