Zhou Jian, Li Hong, Ran Xingwu, Yang Wenying, Li Qiang, Peng Yongde, Li Yanbing, Gao Xin, Luan Xiaojun, Wang Weiqing, Jia Weiping
Department of Endocrinology and Metabolism, Shanghai Jiao Tong University Affiliated Sixth People's Hospital, Shanghai Diabetes Institute, Shanghai Clinical Center for Diabetes, Shanghai, China.
Diabetes Care. 2009 Jul;32(7):1188-93. doi: 10.2337/dc09-0076. Epub 2009 Apr 23.
The widespread clinical application of continuous glucose monitoring (CGM) is limited by the lack of generally accepted reference values. This multicenter study aims to establish preliminary normal reference values for CGM parameters in a sample of healthy Chinese subjects.
A total of 434 healthy individuals with normal glucose regulation completed a 3-day period of glucose monitoring using a CGM system. The 24-h mean blood glucose (24-h MBG) and the percentage of time that subjects' blood glucose levels were >or=140 mg/dl (PT140) and <or=70 mg/dl (PT70) within 24 h were analyzed.
There was excellent compliance of finger stick blood glucose values with CGM measurements for subjects. Among the 434 subjects, the daily blood glucose varied from 76.9 +/- 11.3 to 144.2 +/- 23.2 mg/dl. The 24-h MBG, PT140, and PT70 were 104 +/- 10 mg/dl, 4.1 +/- 5.8%, and 2.4 +/- 5.3%, respectively. As for these parameters, no significant differences were found between men and women. The 95th percentile values were adopted as the upper limits of CGM parameters, which revealed 119 mg/dl (6.6 mmol/l) for 24-h MBG, 17.1% for PT140, and 11.7% for PT70.
We recommend a 24-h MBG value <119 mg/dl, PT140 <17% (4 h), and PT70 <12% (3 h) as normal ranges for the Chinese population.
连续血糖监测(CGM)在临床上的广泛应用受到缺乏普遍认可的参考值的限制。这项多中心研究旨在为健康中国受试者样本建立CGM参数的初步正常参考值。
共有434名血糖调节正常的健康个体使用CGM系统完成了为期3天的血糖监测。分析了24小时平均血糖(24-h MBG)以及受试者血糖水平在24小时内≥140mg/dl(PT140)和≤70mg/dl(PT70)的时间百分比。
受试者的指尖血糖值与CGM测量结果具有良好的一致性。在434名受试者中,每日血糖范围为76.9±11.3至144.2±23.2mg/dl。24-h MBG、PT140和PT70分别为104±10mg/dl、4.1±5.8%和2.4±5.3%。对于这些参数,男性和女性之间未发现显著差异。采用第95百分位数作为CGM参数的上限,结果显示24-h MBG为119mg/dl(6.6mmol/l),PT140为17.1%,PT70为11.7%。
我们建议将24-h MBG值<119mg/dl、PT140<17%(4小时)和PT70<12%(3小时)作为中国人群的正常范围。