Kwan E S, Heilman C B, Shucart W A, Klucznik R P
Department of Radiology, New England Medical Center Hospitals, Boston, Massachusetts.
J Neurosurg. 1991 Dec;75(6):963-8. doi: 10.3171/jns.1991.75.6.0963.
Two patients with distal basilar aneurysms were treated with intra-aneurysmal balloon occlusion. After apparently successful therapy, follow-up angiograms demonstrated aneurysm enlargement with balloon migration distally in the sac. Geometric mismatch between the base of the balloons and the aneurysm neck together with transmitted pulsation through the 2-hydroxyl-ethylmethacrylate (HEMA)-filled balloon directly contributed to aneurysm enlargement. In this report, the authors discuss the problems of progressive aneurysm enlargement due to a "water-hammer effect" and the possibility of hemorrhage following subtotal occlusion.
两名患有基底动脉远端动脉瘤的患者接受了动脉瘤内球囊闭塞治疗。在治疗看似成功后,随访血管造影显示动脉瘤扩大,球囊向瘤腔内远端移位。球囊底部与动脉瘤颈部之间的几何不匹配,以及通过填充甲基丙烯酸2-羟乙酯(HEMA)的球囊传递的搏动,直接导致了动脉瘤扩大。在本报告中,作者讨论了由于“水锤效应”导致动脉瘤逐渐扩大的问题以及次全闭塞后出血的可能性。