Suppr超能文献

关于质子和重带电粒子放射治疗束可实现的临床空间分辨率。

On the clinical spatial resolution achievable with protons and heavier charged particle radiotherapy beams.

作者信息

Andreo Pedro

机构信息

Medical Radiation Physics, Stockholm University and Karolinska University Hospital, SE-171 76 Stockholm, Sweden.

出版信息

Phys Med Biol. 2009 Jun 7;54(11):N205-15. doi: 10.1088/0031-9155/54/11/N01. Epub 2009 May 13.

Abstract

The 'sub-millimetre precision' often claimed to be achievable in protons and light ion beam therapy is analysed using the Monte Carlo code SHIELD-HIT for a broad range of energies. Based on the range of possible values and uncertainties of the mean excitation energy of water and human tissues, as well as of the composition of organs and tissues, it is concluded that precision statements deserve careful reconsideration for treatment planning purposes. It is found that the range of I-values of water stated in ICRU reports 37, 49 and 73 (1984, 1993 and 2005) for the collision stopping power formulae, namely 67 eV, 75 eV and 80 eV, yields a spread of the depth of the Bragg peak of protons and heavier charged particles (carbon ions) of up to 5 or 6 mm, which is also found to be energy dependent due to other energy loss competing interaction mechanisms. The spread is similar in protons and in carbon ions having analogous practical range. Although accurate depth-dose distribution measurements in water can be used at the time of developing empirical dose calculation models, the energy dependence of the spread causes a substantial constraint. In the case of in vivo human tissues, where distribution measurements are not feasible, the problem poses a major limitation. In addition to the spread due to the currently accepted uncertainties of their I-values, a spread of the depth of the Bragg peak due to the varying compositions of soft tissues is also demonstrated, even for cases which could be considered practically identical in clinical practice. For these, the spreads found were similar to those of water or even larger, providing support to international recommendations advising that body-tissue compositions should not be given the standing of physical constants. The results show that it would be necessary to increase the margins of a clinical target volume, even in the case of a water phantom, due to an 'intrinsic basic physics uncertainty', adding to those margins usually considered in normal clinical practice due to anatomical or therapeutic strategy reasons. Individualized patient determination of tissue composition along the complete beam path, rather than CT Hounsfield numbers alone, would also probably be required even to reach 'sub-centimetre precision'.

摘要

使用蒙特卡罗代码SHIELD-HIT,针对广泛的能量范围,分析了质子和轻离子束疗法中常声称可实现的“亚毫米精度”。基于水和人体组织的平均激发能以及器官和组织组成的可能值范围和不确定性,得出结论:出于治疗计划目的,关于精度的表述值得仔细重新审视。研究发现,ICRU报告37、49和73(1984年、1993年和2005年)中给出的用于碰撞阻止本领公式的水的I值范围,即67eV、75eV和80eV,会导致质子和较重带电粒子(碳离子)布拉格峰深度的差异高达5或6mm,并且由于其他能量损失竞争相互作用机制,该差异还与能量有关。质子和具有类似实际射程的碳离子中的差异相似。尽管在开发经验剂量计算模型时可以使用在水中进行的精确深度剂量分布测量,但差异的能量依赖性造成了很大限制。在体内人体组织中,由于无法进行分布测量,该问题构成了主要限制。除了由于目前公认的I值不确定性导致的差异外,还证明了即使在临床实践中可认为实际相同的情况下,软组织组成变化也会导致布拉格峰深度的差异。对于这些情况,发现的差异与水的差异相似甚至更大,这为国际建议提供了支持,即人体组织组成不应被视为物理常数。结果表明,即使对于水模体,由于“内在的基本物理不确定性”,也有必要增加临床靶体积的边界,这在正常临床实践中由于解剖学或治疗策略原因通常会考虑增加边界。甚至为了达到“亚厘米精度”,可能还需要沿整个射线路径对患者的组织组成进行个体化确定,而不仅仅是依靠CT亨氏单位。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验