Medical Imaging Department, Toronto General Hospital, Toronto, Ontario M5G2C4, Canada.
Eur J Radiol. 2010 Aug;75(2):154-8. doi: 10.1016/j.ejrad.2009.04.039. Epub 2009 May 13.
The purpose of this study is to define the relationship between SPECT and CTA measured parameters of left ventricular (LV) function and volumes obtained in a single session using SPECT/64-slice CT hybrid imaging device, and in addition, to assess the reproducibility of LV parameters measured using 64-slice CTA.
Seventy-six patients with suspected or known coronary artery disease underwent cardiac CTA and GSPECT in one session using a hybrid SPECT/CT device. LV end-diastolic volume (EDV), end-systolic volume (ESV) and ejection fraction (EF) were measured on each component of the hybrid device. For the CTA component, these parameters were re-measured by the same investigator and by a second investigator with an interval of 3-54 weeks. Corresponding GSPECT and CTA measured parameters were compared. For CTA, intra-observer and inter-observer variability of LV function and volume measurements were calculated.
A very good correlation was found between the GSPECT and CTA measured LVEF (r=0.81), ESV (r=0.90) and EDV (r=0.82). There was a small positive difference by CTA measured LVEF (3.9+/-14.2%), and more prominent positive differences by CTA measured ESV and EDV (9.8+/-14.8 and 44.9+/-23.1cm(3), respectively). There was excellent reproducibility in the measurements of all parameters with very low intra- and inter-observer variability (r=0.93 for EF and 0.98 for EDV and ESV).
Although a good correlation was found between the EF measurements obtained from CTA and SPECT, interchangeable use of EF measurements between the two modalities should be done cautiously and interchangeable use of LV EDV and ESV should be avoided.
本研究旨在定义使用 SPECT/64 层 CT 混合成像设备在单次检查中测量的左心室(LV)功能和容积的 SPECT 和 CTA 测量参数之间的关系,并评估使用 64 层 CTA 测量的 LV 参数的可重复性。
76 例疑似或已知冠心病患者在一次检查中使用混合 SPECT/CT 设备进行心脏 CTA 和 GSPECT。在混合设备的每个组件上测量 LV 舒张末期容积(EDV)、收缩末期容积(ESV)和射血分数(EF)。对于 CTA 组件,由同一位研究者和第二位研究者在 3-54 周的间隔内重新测量这些参数。比较相应的 GSPECT 和 CTA 测量参数。对于 CTA,计算 LV 功能和容积测量的观察者内和观察者间可变性。
GSPECT 和 CTA 测量的 LVEF(r=0.81)、ESV(r=0.90)和 EDV(r=0.82)之间存在非常好的相关性。CTA 测量的 LVEF(3.9+/-14.2%)有较小的正差异,而 CTA 测量的 ESV 和 EDV 有更明显的正差异(分别为 9.8+/-14.8 和 44.9+/-23.1cm3)。所有参数的测量均具有极好的可重复性,观察者内和观察者间的可变性非常低(EF 的 r=0.93,EDV 和 ESV 的 r=0.98)。
虽然 CTA 和 SPECT 获得的 EF 测量之间存在良好的相关性,但两种模式之间 EF 测量的互换使用应谨慎进行,并且应避免 LV EDV 和 ESV 的互换使用。