Yang R J, Yamada R, Sato M, Sonomura T, Nishida N, Kishi K, Terada M, Maeda M, Yoshioka H, Tsuji K
Department of Radiology, Wakayama Medical College.
Nihon Igaku Hoshasen Gakkai Zasshi. 1991 Aug 25;51(8):970-2.
The authors improved retrievable expandable metallic stent (REMS), instead of the nylon suture, the REMS was connected with the stainless steel thread. Under fluoroscopy the stainless steel thread of the shrinking body could be observed and was strong enough to retract the placed stent, then the stent could be removed by the stainless steel hook. We placed the REMS in the I.V.C. of canine, and removed them after one week placement, we could not remove them after two weeks placement. This study suggested that the REMS was considered to be useful and safe for clinical application.
作者改进了可回收可扩张金属支架(REMS),用不锈钢丝线代替尼龙缝线连接REMS。在荧光透视下可观察到收缩体的不锈钢丝线,其强度足以收回已放置的支架,然后可用不锈钢钩取出支架。我们将REMS放置在犬的下腔静脉中,放置一周后取出,放置两周后无法取出。本研究表明,REMS被认为在临床应用中是有用且安全的。