Lower extremities magnetic resonance angiography with blood pressure cuff compression: quantitative dynamic analysis.

作者信息

Koenigkam-Santos Marcel, Sharma Puneet, Kalb Bobby, Carew John, Oshinski John N, Martin Diego

机构信息

Department of Radiology, Emory University School of Medicine, Atlanta, Georgia, USA.

出版信息

J Magn Reson Imaging. 2009 Jun;29(6):1450-6. doi: 10.1002/jmri.21777.

Abstract

PURPOSE

To quantitatively evaluate changes induced by the application of a femoral blood-pressure cuff (BPC) on run-off magnetic resonance angiography (MRA), which is a method generally previously proposed to reduce venous contamination in the leg.

MATERIALS AND METHODS

This study was Health Insurance Portability and Accountability Act (HIPAA)- and Institutional Review Board (IRB)-compliant. We used time-resolved gradient-echo gadolinium (Gd)-enhanced MRA to measure BPC effects on arterial, venous, and soft-tissue enhancement. Seven healthy volunteers (six men) were studied with the BPC applied at the mid-femoral level unilaterally using a 1.5 T MR system after intravenous injection of Gd-BOPTA. Different statistical tools were used such as the Wilcoxon signed rank test and a cubic smoothing spline fit.

RESULTS

We found that BPC application induces delayed venous filling (as previously described), but also induces significant decreases in arterial inflow, arterial enhancement, vascular-soft tissue contrast, and delayed peak enhancement (which have not been previously measured).

CONCLUSION

The potential benefits from using a BPC for run-off MRA must be balanced against the potential pitfalls, elucidated by our findings.