3T 下使用低剂量钆对比剂的腿部时间分辨磁共振血管成像:初步经验和对比动力学。

Time-resolved MR angiography of the legs at 3 T using a low dose of gadolinium: initial experience and contrast dynamics.

机构信息

Department of Radiology, University Hospitals of Cleveland, Case Western Reserve University, 11100 Euclid Ave, 123 Bolwell Bldg, Cleveland, OH 44106, USA.

出版信息

AJR Am J Roentgenol. 2012 Mar;198(3):686-91. doi: 10.2214/AJR.11.7065.

Abstract

OBJECTIVE

This article describes our initial clinical experience with time-resolved MR angiography (MRA) of the legs using the time-resolved imaging with stochastic trajectories (TWIST) technique with a half dose of gadolinium.

MATERIALS AND METHODS

Thirty-four patients underwent a TWIST examination of the legs at 3 T. Thirty-three patients also underwent a bolus-chase MRA examination in the same setting. Times elapsed between the start of contrast injection and the appearance of contrast material (t(A)) and peak enhancement of the arteries in the legs (t(B)) were analyzed. The number of patients with examinations affected by venous contamination was determined. The differences in t(A) and t(B) between cases in which venous contamination was present or absent were evaluated using a two-tailed Student t test.

RESULTS

The TWIST technique using a half dose of gadolinium provided diagnostic-quality images of all patients. The mean t(A) was 35.5 ± 8.8 (SD) seconds (range, 17.8-60.4 seconds), and the mean t(B) was 59.1 ± 15.1 seconds (range, 31-98.8 seconds). Venous contamination was observed in bolus-chase MRA images of 52.9% of patients. The relationship between venous contamination and t(A) was not statistically significant (p = 0.13). The incidence of venous contamination was higher in patients with lower values of t(B) (p = 0.01).

CONCLUSION

The described low-dose clinical experience with TWIST and the contrast dynamics information gained from this study could aid radiologists in planning protocols for leg MRA examinations.

摘要

目的

本文描述了我们使用随机轨迹时间分辨成像(TWIST)技术和半剂量钆对比剂初步应用于腿部时间分辨磁共振血管成像(MRA)的临床经验。

材料与方法

34 例患者在 3T 磁共振扫描仪上进行 TWIST 腿部扫描。33 例患者还在同一部位进行了团注追踪 MRA 检查。分析了从对比剂注射开始到对比剂出现(t(A))和腿部动脉峰值增强(t(B))的时间。确定受静脉污染影响的检查例数。使用双尾学生 t 检验评估存在或不存在静脉污染时 t(A)和 t(B)的差异。

结果

使用半剂量钆的 TWIST 技术为所有患者提供了诊断质量的图像。平均 t(A)为 35.5 ± 8.8(SD)秒(范围 17.8-60.4 秒),平均 t(B)为 59.1 ± 15.1 秒(范围 31-98.8 秒)。在 52.9%的患者的团注追踪 MRA 图像中观察到静脉污染。静脉污染与 t(A)之间的关系无统计学意义(p = 0.13)。t(B)值较低的患者静脉污染发生率较高(p = 0.01)。

结论

本研究描述的 TWIST 低剂量临床经验以及从这项研究中获得的对比剂动力学信息,可能有助于放射科医生规划腿部 MRA 检查的方案。

相似文献

4
Patient-specific timing for bolus-chase peripheral MR angiography.
J Magn Reson Imaging. 2016 Jan;43(1):249-60. doi: 10.1002/jmri.24968. Epub 2015 Jun 17.

本文引用的文献

7
MR angiography of the lower extremities.下肢磁共振血管造影
AJR Am J Roentgenol. 2008 Jun;190(6):1675-84. doi: 10.2214/AJR.07.2223.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验