Cervinski Mark A, Lockwood Christina M, Ferguson Angela M, Odem Randall R, Stenman Ulf H, Alfthan Henrik, Grenache David G, Gronowski Ann M
Department of Pathology and Immunology, Washington University School of Medicine St. Louis, MO 63110, United States.
Clin Chim Acta. 2009 Aug;406(1-2):81-5. doi: 10.1016/j.cca.2009.05.018. Epub 2009 May 27.
Qualitative point-of-care (POC) tests for human chorionic gonadotropin (hCG) vary in their ability to detect purified hCG variants and there is data to suggest that over-the-counter (OTC) devices might also display similar variability. This could potentially influence the detection of urine hCG in early pregnancy.
Six OTC devices were tested for their ability to detect 5 hCG variants. Ten early pregnancy urine specimens were selected for their diverse expression of hCG variants. The samples were tested with 6 brands of POC and 6 OTC devices.
OTC devices consistently recognized intact hCG, hCGn, and hCGbeta. hCGbetan was consistently recognized by 4 out of 6 brands. One brand inconsistently recognized hCGbetacf. OTC and POC devices varied greatly in their ability to detect hCG in early pregnancy urine, despite the fact that urine samples were adjusted to the same intact hCG concentration. Interestingly, we found that the OTC devices had better analytical sensitivity than the POC devices. Clinitest and First Response demonstrated the lowest hCG detection limits for POC and OTC devices, respectively.
Both OTC and POC devices are capable of detecting hCG concentrations in early pregnancy urine, and OTC devices demonstrated better analytical sensitivity relative to POC devices.
用于检测人绒毛膜促性腺激素(hCG)的定性即时检验(POC)在检测纯化的hCG变体的能力方面存在差异,并且有数据表明非处方(OTC)设备可能也存在类似的变异性。这可能会潜在地影响早期妊娠尿液中hCG的检测。
测试了六种OTC设备检测5种hCG变体的能力。选择了10份早期妊娠尿液标本,因其hCG变体表达多样。用6个品牌的POC和6种OTC设备对样本进行检测。
OTC设备始终能识别完整的hCG、hCGn和hCGβ。6个品牌中有4个品牌始终能识别hCGβn。有一个品牌对hCGβcf的识别不稳定。尽管将尿液样本调整到相同的完整hCG浓度,但OTC和POC设备在检测早期妊娠尿液中hCG的能力上差异很大。有趣的是,我们发现OTC设备的分析灵敏度比POC设备更好。Clinitest和First Response分别展示了POC和OTC设备的最低hCG检测限。
OTC和POC设备都能够检测早期妊娠尿液中的hCG浓度,并且相对于POC设备,OTC设备展示出了更好的分析灵敏度。