Suppr超能文献

基于便携式伽马相机的孤立肢体灌注监测系统。

Monitoring system for isolated limb perfusion based on a portable gamma camera.

作者信息

Orero A, Vidal-Sicart S, Roé N, Muxí A, Rubí S, Duch J, Rull R, Pavón N, Pons F, Pavía J

机构信息

Servei de Medicina Nuclear, Hospital Clinic, Barcelona, Spain.

出版信息

Nuklearmedizin. 2009;48(4):166-72. doi: 10.3413/nukmed-0223. Epub 2009 Jun 2.

Abstract

BACKGROUND

The treatment of malignant melanoma or sarcomas on a limb using extremity perfusion with tumour necrosis factor (TNF-alpha) and melphalan can result in a high degree of systemic toxicity if there is any leakage from the isolated blood territory of the limb into the systemic vascular territory. Leakage is currently controlled by using radiotracers and heavy external probes in a procedure that requires continuous manual calculations. The aim of this work was to develop a light, easily transportable system to monitor limb perfusion leakage by controlling systemic blood pool radioactivity with a portable gamma camera adapted for intraoperative use as an external probe, and to initiate its application in the treatment of MM patients.

METHODS

A special collimator was built for maximal sensitivity. Software for acquisition and data processing in real time was developed. After testing the adequacy of the system, it was used to monitor limb perfusion leakage in 16 patients with malignant melanoma to be treated with perfusion of TNF-alpha and melphalan.

RESULTS

The field of view of the detector system was 13.8 cm, which is appropriate for the monitoring, since the area to be controlled was the precordial zone. The sensitivity of the system was 257 cps/MBq. When the percentage of leakage reaches 10% the associated absolute error is +/-1%. After a mean follow-up period of 12 months, no patients have shown any significant or lasting side-effects. Partial or complete remission of lesions was seen in 9 out of 16 patients (56%) after HILP with TNF-alpha and melphalan.

CONCLUSION

The detector system together with specially developed software provides a suitable automatic continuous monitoring system of any leakage that may occur during limb perfusion. This technique has been successfully implemented in patients for whom perfusion with TNF-alpha and melphalan has been indicated.

摘要

背景

使用肿瘤坏死因子(TNF-α)和美法仑进行肢体灌注治疗肢体恶性黑色素瘤或肉瘤时,如果肢体隔离血域有任何渗漏进入体循环血管域,可能会导致高度的全身毒性。目前通过在一个需要持续人工计算的过程中使用放射性示踪剂和重型外部探头来控制渗漏。这项工作的目的是开发一种轻便、易于运输的系统,通过使用适合术中用作外部探头的便携式γ相机控制体循环血池放射性来监测肢体灌注渗漏,并将其应用于黑色素瘤患者的治疗。

方法

制造了一个用于最大灵敏度的特殊准直器。开发了用于实时采集和数据处理的软件。在测试系统的适用性后,将其用于监测16例拟用TNF-α和美法仑灌注治疗的恶性黑色素瘤患者的肢体灌注渗漏情况。

结果

探测器系统的视野为13.8厘米,适合监测,因为要控制的区域是心前区。系统的灵敏度为257 cps/MBq。当渗漏百分比达到10%时,相关的绝对误差为±1%。平均随访12个月后,没有患者出现任何明显或持久的副作用。16例患者中有9例(56%)在接受TNF-α和美法仑的热灌注隔离肢体化疗(HILP)后出现病变部分或完全缓解。

结论

探测器系统与专门开发的软件一起提供了一个合适的自动连续监测系统,用于监测肢体灌注过程中可能发生的任何渗漏。这项技术已成功应用于适合使用TNF-α和美法仑进行灌注治疗的患者。