J Refract Surg. 2009 May;25(5):429-34. doi: 10.3928/1081597X-20090422-05.

Abstract

PURPOSE

To compare measurements obtained with the VISX WaveScan and NIDEK OPD-Scan (ARK 10000).

METHODS

The VISX WaveScan is a wavefront aberrometer that uses Hartmann-Shack technology. The NIDEK OPD-Scan uses time-based retinoscopy (dynamic skiascopy). Measurements taken of 92 eyes from 51 patients using both aberrometers were retrieved and compared. Measurements included sphere, cylinder, axis, and Zernike coefficients of second to sixth radial order aberrations. The Zernike coefficients were normalized to a 6.0-mm pupil and used to calculate the root-mean-square (RMS) values of the radial orders, angular orders, total higher order aberrations, third order trefoil, third order coma, total spherical aberration, and total coma. Paired t tests were used to compare differences between measurements.

RESULTS

Although no significant difference was found between the NIDEK OPD-Scan and subjective manifest refraction in sphere and axis, the NIDEK OPD-Scan measured significantly higher cylinder. No significant difference was found between the VISX WaveScan and subjective manifest refraction. The NIDEK OPD-Scan had significantly lower total higher order, second radial order, third order coma, total spherical aberration, total coma, and second angular order RMS values. The VISX WaveScan had significantly lower cylinder, trefoil, third-angular order, and fifth-angular order RMS values.

CONCLUSIONS

Specific and significant differences were noted between the measurements taken by the VISX WaveScan and NIDEK OPD-Scan wavefront aberrometers.

摘要

目的

比较VISX WaveScan和尼德克OPD-Scan(ARK 10000)所获得的测量结果。

方法

VISX WaveScan是一种使用哈特曼-夏克技术的波前像差仪。尼德克OPD-Scan采用基于时间的检影法(动态视网膜检影法)。检索并比较了使用这两种像差仪对51例患者的92只眼睛进行的测量结果。测量内容包括球镜、柱镜、轴位以及二阶至六阶径向像差的泽尼克系数。将泽尼克系数归一化为6.0毫米的瞳孔直径,并用于计算径向阶数、角向阶数、总高阶像差、三阶三叶形像差、三阶彗差、总球差和总彗差的均方根(RMS)值。采用配对t检验比较测量结果之间的差异。

结果

尽管尼德克OPD-Scan与主观显验光在球镜和轴位上未发现显著差异,但尼德克OPD-Scan测量的柱镜度数显著更高。VISX WaveScan与主观显验光之间未发现显著差异。尼德克OPD-Scan的总高阶、二阶径向阶数、三阶彗差、总球差、总彗差和二阶角向阶数的RMS值显著更低。VISX WaveScan的柱镜、三叶形像差、三阶角向阶数和五阶角向阶数的RMS值显著更低。

结论

在VISX WaveScan和尼德克OPD-Scan波前像差仪所进行的测量之间发现了特定且显著的差异。