Bogue Conor O, John Philip R, Connolly Bairbre L, Rea David J, Amaral Joao G
Division of Image Guided Therapy, Department of Diagnostic Imaging, The Hospital for Sick Children and University of Toronto, Toronto M5G 1X8, ON, Canada.
Pediatr Radiol. 2009 Oct;39(10):1110-3. doi: 10.1007/s00247-009-1340-4. Epub 2009 Jul 9.
We report a case of penetration of the inferior vena cava (IVC) by all four primary struts of a Celect caval filter in a 17-year-old girl with Klippel-Trénaunay syndrome. The girl presented with acute lower abdominal and right leg pain 17 days after filter insertion. An abdominal radiograph demonstrated that the filter had moved caudally and that the primary struts had splayed considerably since insertion. Contrast-enhanced CT confirmed that all four primary struts had penetrated the IVC wall. There was a small amount of retroperitoneal hemorrhage. The surrounding vessels and viscera were intact. The filter was subsequently retrieved without complication.
我们报告了一例17岁患有克-特综合征(Klippel-Trénaunay syndrome)的女孩,其Celect下腔静脉滤器的所有四个主要支柱穿透了下腔静脉(IVC)。该女孩在滤器植入后17天出现急性下腹部和右腿疼痛。腹部X线片显示滤器已向尾侧移动,自植入后主要支柱已明显张开。增强CT证实所有四个主要支柱均穿透了下腔静脉壁。有少量腹膜后出血。周围血管和内脏完好无损。随后顺利取出滤器,未出现并发症。