To evaluate Orbscan II measurements without contact lens (CL) removal to monitor in vivo posterior corneal curvature in CL (hypoxia)-induced corneal swelling during extended wear (EW).
Orbscan II corneal topographer was used to measure posterior corneal curvature 1 week before CL wearing and during 1 week of EW. Central corneal thickness (CCT) was also measured during the 2-week study with OCT to determine the corneal swelling percentage. High Dk (lotrafilcon A) and low Dk (etafilcon A) soft CL were randomly fitted in EW in the right and left eyes of 20 subjects with normal ocular health. Orbscan and OCT were also performed without CL removal after 3 and 7 days of wear.
Low Dk lenses induced significantly higher corneal swelling (2.41+/-2.65%), as compared to the high Dk lenses (0.17+/-2.28%; P<0.05 repeated measures ANOVA (Re-ANOVA and paired t-test) measures with OCT. The low Dk lens induced a significantly posterior corneal flattening compared to the high Dk lens (0.036+/-0.06mm versus 0.006+/-0.04mm; P<0.05 Re-ANOVA and paired t-test). High repeatability for Orbscan-measured posterior best fit sphere (BFS) radii with (r(2)=0.15, 95% limits of agreement (LoA)=0.10 to -0.06) and without (r(2)=0.00, 95% LoA=0.08 to -0.08) CL removal was also observed and suggest that posterior corneal changes was not due to Orbscan variability.
Orbscan is a non-invasive and reproducible technique to evaluate and monitoring posterior corneal curvature in vivo in CL (hypoxia)-induced corneal swelling during EW.
评估 Orbscan II 测量值,无需去除隐形眼镜 (CL),以监测延长佩戴期间 CL(缺氧)诱导的角膜肿胀的活体后角膜曲率。
在佩戴 CL 前一周和延长佩戴的一周内,使用 Orbscan II 角膜地形图仪测量后角膜曲率。在 2 周的研究中,还使用 OCT 测量中央角膜厚度 (CCT),以确定角膜肿胀百分比。20 名眼部健康正常的受试者的右眼和左眼随机佩戴高 Dk(乐透福龙 A)和低 Dk(依视福龙 A)软 CL 进行 EW。还在佩戴后 3 天和 7 天无需去除 CL 的情况下进行 Orbscan 和 OCT 检查。
低 Dk 镜片引起的角膜肿胀明显更高(2.41+/-2.65%),而高 Dk 镜片则为(0.17+/-2.28%);(P<0.05 重复测量方差分析(Re-ANOVA 和配对 t 检验)与 OCT 测量值。低 Dk 镜片引起的后角膜变平明显大于高 Dk 镜片(0.036+/-0.06mm 比 0.006+/-0.04mm;P<0.05 Re-ANOVA 和配对 t 检验)。使用 Orbscan 测量的后最佳拟合球(BFS)半径的重复性也很高,无论是否去除 CL(r(2)=0.15,95%置信区间(LoA)=0.10 至 -0.06),r(2)=0.00,95% LoA=0.08 至 -0.08),这表明后角膜变化不是由于 Orbscan 变异性引起的。
Orbscan 是一种非侵入性和可重复的技术,可用于评估和监测延长佩戴期间 CL(缺氧)诱导的角膜肿胀的活体后角膜曲率。