Bretschneider Thorsten, Koop Urte, Schreiner Volker, Wenck Horst, Jaspers Sören
Beiersdorf Research & Development, Beiersdorf AG, Hamburg, Germany.
Skin Res Technol. 2009 Aug;15(3):364-9. doi: 10.1111/j.1600-0846.2009.00374.x.
Currently, the body scanner, using laser-triangulation, is one of the most precise measuring tools for the rapid quantification of body shape. The VITUS body scanner is a laser-based system based on a principle called triangulation and the scan produced describes the distance to a surface at each point in the picture. The body scanner has multiple applications such as determining body measurements for tailoring, anthropometric investigations and cosmetic surgery. There are also intensive investigations into the effect of weight gain and thus body shape on health risks. In order to be of value, the body scanner needs to generate precise, accurate and reproducible data.
To determine the precision and reproducibility of the VITUS XXL 3D body scanner.
The measurements of geometric shapes (cones, columns) and human body parts (mid-thigh) were compared using a measuring tape and the body scanner.
The precision of the measurements of the circumferences of a truncated cone and a column was within 1 mm of the actual values (0.29%). The reproducibility of the measurements was very good. The standard deviation in the measurement of a truncated cone was only 0.13% of the actual value. Likewise, the standard deviation of the thigh measurement of 12 human subjects was <1%.
These results show that the body scanner can accurately, precisely and reproducibly measure the circumference of objects and human body parts.
目前,采用激光三角测量法的人体扫描仪是快速量化体型最精确的测量工具之一。VITUS人体扫描仪是一种基于激光的系统,其原理是三角测量法,生成的扫描图像描述了图片中每个点到表面的距离。人体扫描仪有多种应用,如为裁剪确定身体尺寸、人体测量研究和整容手术。人们还深入研究了体重增加以及体型对健康风险的影响。为了具有价值,人体扫描仪需要生成精确、准确且可重复的数据。
确定VITUS XXL 3D人体扫描仪的精度和可重复性。
使用卷尺和人体扫描仪对几何形状(圆锥体、圆柱体)和人体部位(大腿中部)进行测量并比较。
截头圆锥体和圆柱体周长测量的精度在实际值的1毫米范围内(0.29%)。测量的可重复性非常好。截头圆锥体测量的标准差仅为实际值的0.13%。同样,12名受试者大腿测量的标准差<1%。
这些结果表明,人体扫描仪能够准确、精确且可重复地测量物体和人体部位的周长。