Regional anaesthesia elastomeric pump performance after a single use and subsequent refill: a laboratory study.

作者信息

Grant C R K, Fredrickson M J

机构信息

Ninewells Hospital and Medical School, Dundee, UK.

出版信息

Anaesthesia. 2009 Jul;64(7):770-5. doi: 10.1111/j.1365-2044.2009.05941.x.

Abstract

Ambulatory local anaesthetic delivery systems are often limited by a short effective duration of infusion. Prolonging nerve blockade by substituting a new pump as recommended by the manufacturers, represents a substantial consumable item cost ($US 300-500). We therefore evaluated the flow delivery performance of 31 single model elastomeric devices (all with a 2 ml.h(-1) background and 5 ml every hour bolus capability) that had been filled, used in clinical practice and then refilled in the laboratory. For the second infusion, there was a pattern of over-infusion (< 10 ml.h(-1)) in the first hour; however, all pumps depreciated to < 150% of predicted by the second hour. The subsequent performance of all pumps was not only within safe limits, but also predominantly within the range (background infusion +/- 15%, bolus +10/-20%) specified by the manufacturer for primary infusion. We conclude that this elastomeric regional anaesthesia pump design performs satisfactorily after having been refilled following a single previous use.

摘要

门诊局部麻醉给药系统通常受输注有效持续时间短的限制。按照制造商的建议更换新泵来延长神经阻滞时间,这意味着可观的耗材成本(300 - 500美元)。因此,我们评估了31个单一型号的弹性体装置的流量输送性能(所有装置均具有2 ml.h⁻¹的背景输注速率和每小时5 ml的推注能力),这些装置已在临床实践中使用并填充,然后在实验室中重新填充。在第二次输注时,第一个小时出现了输注过量(< 10 ml.h⁻¹)的情况;然而,到第二个小时,所有泵的输注量均降至预测值的150%以下。所有泵随后的性能不仅在安全范围内,而且主要在制造商规定的初次输注范围(背景输注±15%,推注+10/-20%)内。我们得出结论,这种弹性体区域麻醉泵设计在单次使用后重新填充后表现令人满意。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验