尿路病原体与具有不同表面特性的泌尿外科支架表面的相互作用。

Uropathogen interaction with the surface of urological stents using different surface properties.

作者信息

Lange Dirk, Elwood Chelsea N, Choi Kenny, Hendlin Kari, Monga Manoj, Chew Ben H

机构信息

Stone Centre at Vancouver General Hospital, Department of Urologic Sciences, University of British Columbia, Vancouver, Canada.

出版信息

J Urol. 2009 Sep;182(3):1194-200. doi: 10.1016/j.juro.2009.05.008. Epub 2009 Jul 21.

Abstract

PURPOSE

Ureteral stents commonly become infected or encrusted. Various coatings have been developed to decrease bacterial adherence. To our knowledge there has been no in vitro testing of coating with heparin to date. We determined the effects of heparin coating on bacterial adherence of common uropathogens and physical stent properties.

MATERIALS AND METHODS

Heparin coated Radiance ureteral stents (Cook) and noncoated Endo-Sof control stents were tested against triclosan eluting Triumph(R) stents and noneluting Polaris control stents for adherence of Escherichia coli, Klebsiella pneumoniae, Enterococcus faecalis, Staphylococcus aureus and Pseudomonas aeruginosa for 7 days. Adherent bacteria were determined and biofilms were visualized using fluorescent dyes. Radial, tensile and coil strength of the Radiance and Polaris stents was compared to determine the effect of heparin coating on physical stent characteristics.

RESULTS

Heparin coating did not decrease bacterial adhesion compared to its control. E. coli adhesion was limited by all stents tested. The Polaris stent showed significantly greater resistance to bacterial adherence for Klebsiella, Pseudomonas and Enterococcus than the Endo-Sof and Radiance stents but was more susceptible to S. aureus adherence. The Triumph stent resisted all bacteria except Pseudomonas and Enterococcus. Mature biofilms were observed on all stents with lower viability on the Triumph stent. Radiance stents showed higher tensile and lower compression strength than its control.

CONCLUSIONS

Heparin coating does not decrease bacterial adherence to ureteral stents. Drug eluting antimicrobials have an inhibitory effect on bacterial adherence and the Polaris stent showed the least bacterial adherence of the nondrug eluting ureteral stents tested.

摘要

目的

输尿管支架常发生感染或结痂。已研发出多种涂层以减少细菌黏附。据我们所知,迄今为止尚未对肝素涂层进行体外测试。我们确定了肝素涂层对常见尿路病原体细菌黏附及支架物理性能的影响。

材料与方法

将肝素涂层的Radiance输尿管支架(库克公司)和未涂层的Endo-Sof对照支架与三氯生洗脱的Triumph支架和非洗脱的Polaris对照支架进行测试,观察大肠杆菌、肺炎克雷伯菌、粪肠球菌、金黄色葡萄球菌和铜绿假单胞菌在7天内的黏附情况。测定黏附细菌数量,并使用荧光染料观察生物膜。比较Radiance和Polaris支架的径向、拉伸和卷曲强度,以确定肝素涂层对支架物理特性的影响。

结果

与对照相比,肝素涂层并未减少细菌黏附。所有测试支架对大肠杆菌的黏附均有抑制作用。Polaris支架对克雷伯菌、假单胞菌和肠球菌的细菌黏附抵抗力明显高于Endo-Sof和Radiance支架,但对金黄色葡萄球菌黏附更敏感。Triumph支架对除假单胞菌和肠球菌外的所有细菌均有抵抗力。在所有支架上均观察到成熟生物膜,Triumph支架上的生物膜活力较低。Radiance支架的拉伸强度较高,压缩强度较低。

结论

肝素涂层不能减少细菌对输尿管支架的黏附。药物洗脱抗菌剂对细菌黏附有抑制作用,Polaris支架在所测试的非药物洗脱输尿管支架中细菌黏附最少。