Sanfelippo Michael J, Zinsmaster Wendy, Scherr Doris L, Shaw Gene R
Coagulation Services, Marshfield Labs, Marshfield Clinic, 1000 North Oak Avenue, Marshfield, WI 54449, USA.
Clin Med Res. 2009 Sep;7(3):103-5. doi: 10.3121/cmr.2009.836. Epub 2009 Jul 22.
A warfarin treated patient unexpectedly presented with an elevated international normalized ratio (INR). Repeat testing in two laboratories gave conflicting results. The chromogenic assay of factor X was used to determine the correct INR result. The patient had laboratory results consistent with a dysfibrinogenemia, which prevented detection of the endpoint with a photo-optical detection system. The chromogenic assay of factor X is recommended for monitoring patients on warfarin when the INR cannot be accurately determined due to interference with the fibrin endpoint in the INR.