Sensitivity considerations for large molecule detection by capillary electrophoresis-electrospray ionization mass spectrometry.

作者信息

Smith R D, Loo J A, Edmonds C G, Barinaga C J, Udseth H R

机构信息

Chemical Sciences Department, Pacific Northwest Laboratories, Richland, WA 99352.

出版信息

J Chromatogr. 1990 Sep 7;516(1):157-65. doi: 10.1016/s0021-9673(01)90213-4.

Abstract

The use of the electrospray ionization (ESI) method for interfacing capillary electrophoresis with mass spectrometry (CE-MS) is particularly well suited for the analysis of large molecules due to the multiple charging phenomenon. While ionization efficiency is very high, the available ion current is dispersed over more peaks so that the maximum peak intensity obtainable declines significantly for large molecules. Sensitivity with ESI can be improved by operation at very low flow-rates, an ideal situation for CE-MS. These and other considerations related to sensitivity are illustrated using ESI-MS measurements for cytochrome c.

摘要

由于多电荷现象,电喷雾电离(ESI)方法用于毛细管电泳与质谱联用(CE-MS)特别适合于大分子分析。虽然电离效率非常高,但可用离子电流分散在更多峰上,因此对于大分子可获得的最大峰强度会显著下降。通过在非常低的流速下操作可以提高ESI的灵敏度,这对于CE-MS来说是理想的情况。使用细胞色素c的ESI-MS测量说明了这些以及与灵敏度相关的其他考虑因素。