IEEE Trans Med Imaging. 2009 Dec;28(12):2007-14. doi: 10.1109/TMI.2009.2027436. Epub 2009 Jul 21.
The performance of a position sensitive gamma-ray scintillator detector (PSD) is described. This PSD is based on a lutetium yttrium oxyorthosilicate (LYSO) crystal read out by a crossed-wire anode position sensitive photomultiplier tube (PSPMT). The main difference with respect to similar existing devices is the individual multi-anode readout (IMAR) approach that is followed here. This method allows to exploit better the intrinsic characteristics of the PSPMT, thus yielding better linearity, improved spatial resolution, and a larger field of view. The new detector is intended for the characterization of 3-D position sensitive germanium detectors.
介绍了一种位置灵敏γ射线闪烁探测器(PSD)的性能。该 PSD 基于由交叉线阳极位置灵敏光电倍增管(PSPMT)读取的硅酸镥钇(LYSO)晶体。与类似的现有设备相比,主要区别在于这里采用的是独立多阳极读出(IMAR)方法。这种方法可以更好地利用 PSPMT 的固有特性,从而获得更好的线性度、改进的空间分辨率和更大的视场。新型探测器旨在用于 3D 位置灵敏锗探测器的特性描述。