Jacob Sony, Cherian Prasad K, Ghumman Waqas S, Das Mithilesh K
Division of Cardiology/Electrophysiology, Wayne State University, Detroit, MI, USA.
J Interv Card Electrophysiol. 2010 Sep;28(3):221-5. doi: 10.1007/s10840-009-9415-6. Epub 2009 Jul 28.
Patients implanted with left ventricular assist devices (LVAD) may have implantable cardioverter defibrillators (ICD) implanted for sudden cardiac death prevention. This opens the possibility of device-device communication interactions and thus interferences. We present a case of such interaction that led to ICD communication failure following the activation of an LVAD. In this paper, we describe a practical solution to circumvent the communication interference and review the communication links of ICDs and possible mechanisms of ICD-LVAD interactions.
植入左心室辅助装置(LVAD)的患者可能会植入植入式心脏复律除颤器(ICD)以预防心脏性猝死。这就开启了设备与设备之间通信交互以及干扰的可能性。我们呈现了这样一个因LVAD激活导致ICD通信失败的交互案例。在本文中,我们描述了一种规避通信干扰的实用解决方案,并回顾了ICD的通信链路以及ICD与LVAD相互作用的可能机制。