A stability-indicating HPLC assay with diode array detection for the determination of a benzylpenicillin prodrug in aqueous solutions.

机构信息

School of Pharmacy, University of Otago, P.O. Box 56, Dunedin, New Zealand.

出版信息

J Pharm Biomed Anal. 2009 Dec 5;50(5):841-6. doi: 10.1016/j.jpba.2009.06.046. Epub 2009 Jul 3.

Abstract

The aim of this study was to develop a stability-indicating HPLC assay for the determination of penethamate (PNT), an ester prodrug of benzylpenicillin (BP), in aqueous solutions. The method was validated by subjecting PNT to forced decomposition under stress conditions of acid, alkali, water hydrolysis and oxidation. A quenching solution was developed to limit degradation to negligible levels before and during the analysis. Both PNT and BP were simultaneously determined and separated in presence of degradation products on a C(18) column using a mobile phase consisting of methanol-acetonitrile-acetate buffer. Different degradation products were formed in the stress conditions. The peak purity indexes of PNT and BP obtained by diode array detection were >0.999, confirming the absence of other co-eluting substances. The assay was linear for both analytes in the concentration range 1-100 microg mL(-1). The LOD and LOQ of PNT were 0.03 and 0.09 microg mL(-1) respectively. Degradation of PNT followed pseudo-first-order kinetics with t(1/2) of 43.6 min at pH 2.01 and 4.2 min at pH 9.31. In addition, the absence of BP in the acidic solutions of PNT emphasises the futility of monitoring BP to assess the stability of PNT. In conclusion, the assay is rapid and stability-indicating with adequate precision and accuracy, and in conjunction with the quenching solution, can be used for stability studies of PNT with simultaneous quantitation of BP. The degradation studies provide useful information for formulation development of PNT.

摘要

本研究旨在开发一种适用于测定苯芐青霉素(BP)酯前体药物喷他脒(PNT)在水溶液中稳定性的 HPLC 分析方法。该方法通过在酸、碱、水水解和氧化等强制降解条件下对 PNT 进行了验证。开发了一种猝灭溶液,以在分析之前和分析过程中将降解限制在可忽略的水平。在存在降解产物的情况下,PNT 和 BP 同时在 C(18)柱上用甲醇-乙腈-醋酸缓冲液作为流动相进行同时测定和分离。在应激条件下形成了不同的降解产物。二极管阵列检测得到的 PNT 和 BP 的峰纯度指数均>0.999,证实不存在其他共洗脱物质。该分析方法对于 PNT 和 BP 的浓度范围为 1-100μg mL(-1) 均呈线性。PNT 的 LOD 和 LOQ 分别为 0.03 和 0.09μg mL(-1)。PNT 在 pH 2.01 时的降解遵循伪一级动力学,半衰期为 43.6 分钟,在 pH 9.31 时为 4.2 分钟。此外,在 PNT 的酸性溶液中不存在 BP,强调了监测 BP 以评估 PNT 的稳定性是徒劳的。总之,该分析方法快速、具有稳定性指示性,具有足够的精密度和准确性,并且与猝灭溶液结合使用,可以用于 PNT 的稳定性研究,同时定量测定 BP。降解研究为 PNT 的制剂开发提供了有用的信息。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验