Malakova Jana, Pavek Petr, Svecova Lucie, Jokesova Iveta, Zivny Pavel, Palicka Vladimir
Institute of Clinical Biochemistry and Diagnostics, Charles University in Prague, Faculty of Medicine and University Hospital Hradec Králové, Sokolská 581, CZ-50005 Hradec Králové, Czech Republic.
J Chromatogr B Analyt Technol Biomed Life Sci. 2009 Oct 1;877(27):3226-30. doi: 10.1016/j.jchromb.2009.07.009. Epub 2009 Jul 15.
Warfarin is a well-known anticoagulant agent that occurs in two enantiomers, (R)-(+)-warfarin and (S)-(-)-warfarin. A new liquid chromatography method for the determination of both enantiomers was developed, validated and applied in in vitro studies with the aim of evaluating the accumulation of (R)-warfarin and (S)-warfarin in the hepatoma HepG2 cell line. OptiMEM cell cultivation medium samples and cellular lysates were purified using Waters Oasis MAX extraction cartridges. The chiral separation of warfarin and the internal standard p-chlorowarfarin enantiomers was performed on an Astec Chirobiotic V2 column at a flow rate of 1.2mL/min. The mobile phase was composed of 31% acetonitrile, 5% of methanol and 64% of ammonium acetate buffer (10mmol/L, pH 4.1). The enantiomers were quantified using a fluorescence detector (lambda(excit)=320nm, lambda(emiss)=415nm). The limit of detection was found to be 0.121micromol/L of (S)-warfarin and 0.109micromol/L of (R)-warfarin. The range of applicability and linearity was estimated from 0.25 to 100micromol/L. The precision ranged from 1.3% to 12.2% of the relative standard deviation, and the accuracy reached acceptable values from 95.5% to 108.4%. The new bioanalytical method confirmed the same accumulation of (R)-warfarin and (S)-warfarin in the hepatoma HepG2 cell line.
华法林是一种著名的抗凝剂,它有两种对映体,即(R)-(+)-华法林和(S)-(-)-华法林。开发并验证了一种用于测定这两种对映体的新型液相色谱方法,并将其应用于体外研究,目的是评估(R)-华法林和(S)-华法林在肝癌HepG2细胞系中的积累情况。使用Waters Oasis MAX萃取柱对OptiMEM细胞培养基样品和细胞裂解物进行纯化。在Astec Chirobiotic V2柱上以1.2mL/min的流速进行华法林和内标对氯华法林对映体的手性分离。流动相由31%乙腈、5%甲醇和64%醋酸铵缓冲液(10mmol/L,pH 4.1)组成。使用荧光检测器(激发波长λ(excit)=320nm,发射波长λ(emiss)=415nm)对映体进行定量。检测限为(S)-华法林0.121μmol/L,(R)-华法林0.109μmol/L。适用范围和线性范围估计为0.25至100μmol/L。精密度范围为相对标准偏差的1.3%至12.2%,准确度达到95.5%至108.4%的可接受值。新的生物分析方法证实了(R)-华法林和(S)-华法林在肝癌HepG2细胞系中的积累情况相同。