Kyriakou Yiannis, Prell Daniel, Kalender Willi A
Institute of Medical Physics, University of Erlangen-Nürnberg, Germany.
Phys Med Biol. 2009 Sep 7;54(17):N385-91. doi: 10.1088/0031-9155/54/17/N02. Epub 2009 Aug 6.
In high-resolution micro CT using flat detectors (FD), imperfect or defect detector elements may cause concentric-ring artifacts due to their continuous over- or underestimation of attenuation values, which often disturb image quality. We here present a dedicated image-based ring artifact correction method for high-resolution micro CT, based on median filtering of the reconstructed image and working on a transformed version of the reconstructed images in polar coordinates. This post-processing method reduced ring artifacts in the reconstructed images and improved image quality for phantom and in in vivo scans. Noise and artifacts were reduced both in transversal and in multi-planar reformations along the longitudinal axis.
在使用平板探测器(FD)的高分辨率微型计算机断层扫描(micro CT)中,不完善或有缺陷的探测器元件可能因其对衰减值的持续高估或低估而导致同心环伪影,这常常干扰图像质量。我们在此提出一种专门用于高分辨率微型计算机断层扫描的基于图像的环形伪影校正方法,该方法基于对重建图像进行中值滤波,并在极坐标下对重建图像的变换版本进行处理。这种后处理方法减少了重建图像中的环形伪影,并提高了体模和活体扫描的图像质量。在沿纵轴的横向和多平面重建中,噪声和伪影均有所减少。