In vivo dosimetry with a linear MOSFET array to evaluate the urethra dose during permanent implant brachytherapy using iodine-125.

作者信息

Bloemen-van Gurp Esther J, Haanstra Björk K C, Murrer Lars H P, van Gils Francis C J M, Dekker Andre L A J, Mijnheer Ben J, Lambin Philippe

机构信息

Department of Radiation Oncology (MAASTRO), GROW, University Hospital Maastricht, the Netherlands.

出版信息

Int J Radiat Oncol Biol Phys. 2009 Nov 15;75(4):1266-72. doi: 10.1016/j.ijrobp.2009.04.042. Epub 2009 Aug 6.

Abstract

PURPOSE

To develop a technique to monitor the dose rate in the urethra during permanent implant brachytherapy using a linear MOSFET array, with sufficient accuracy and without significantly extending the implantation time.

METHODS AND MATERIALS

Phantom measurements were performed to determine the optimal conditions for clinical measurements. In vivo measurements were performed in 5 patients during the (125)I brachytherapy implant procedure. To evaluate if the urethra dose obtained in the operating room with the ultrasound transducer in the rectum and the patient in treatment position is a reference for the total accumulated dose; additional measurements were performed after the implantation procedure, in the recovery room.

RESULTS

In vivo measurements during and after the implantation procedure agree very well, illustrating that the ultrasound transducer in the rectum and patient positioning do not influence the measured dose in the urethra. In vivo dose values obtained during the implantation are therefore representative for the total accumulated dose in the urethra. In 5 patients, the dose rates during and after the implantation were below the maximum dose rate of the urethra, using the planned seed distribution.

CONCLUSION

In vivo dosimetry during the implantation, using a MOSFET array, is a feasible technique to evaluate the dose in the urethra during the implantation of (125)I seeds for prostate brachytherapy.

摘要

目的

开发一种技术,使用线性MOSFET阵列在永久性植入近距离放射治疗期间监测尿道中的剂量率,具有足够的准确性且不会显著延长植入时间。

方法和材料

进行模体测量以确定临床测量的最佳条件。在5例患者的(125)I近距离放射治疗植入过程中进行了体内测量。为了评估在手术室中,将超声换能器置于直肠且患者处于治疗体位时所获得的尿道剂量是否可作为总累积剂量的参考;在植入手术后,在恢复室进行了额外的测量。

结果

植入过程中和植入后的体内测量结果非常吻合,表明直肠中的超声换能器和患者体位不会影响尿道中的测量剂量。因此,植入过程中获得的体内剂量值代表了尿道中的总累积剂量。在5例患者中,使用计划的种子分布,植入过程中和植入后的剂量率均低于尿道的最大剂量率。

结论

在植入(125)I种子进行前列腺近距离放射治疗期间,使用MOSFET阵列进行体内剂量测定是一种可行的技术,可用于评估尿道中的剂量。