Kaiser Patricia, Akerboom Theodorus, Dux László, Reinauer Hans
Reference Measurement Laboratory, Instand e.V., Düsseldorf, Germany.
Ger Med Sci. 2006 Sep 21;4:Doc06.
The reference measurement procedure for determination of HbA(1c) (glycated haemoglobin) using HPLC(high performance liquid chromatography)-ESI(electrospray ionisation)-MS(mass spectrometry) has been modified. Main modifications were a change in the buffer composition of the HPLC, a change in the gradient elution profile and the introduction of a post-column splitting system. The long-term stability of the HPLC-ESI-MS system proved to be of high importance to get reproducible results. Using the modified HPLC conditions stable retention times, high peak symmetry and a good column batch-to-batch reproducibility of the analytical results were obtained. The mean deviation from the target values was less than 0.7% with an imprecision of less than 2%.
采用高效液相色谱(HPLC)-电喷雾电离(ESI)-质谱(MS)法测定糖化血红蛋白(HbA(1c))的参考测量程序已得到改进。主要改进包括HPLC缓冲液成分的改变、梯度洗脱曲线的变化以及柱后分流系统的引入。事实证明,HPLC-ESI-MS系统的长期稳定性对于获得可重复的结果至关重要。采用改进后的HPLC条件,可获得稳定的保留时间、较高的峰对称性以及各批次色谱柱之间良好的分析结果重现性。与目标值的平均偏差小于0.7%,不精密度小于2%。