可诱导的小梁金属胫骨单体部件的位移。

Inducible displacement of a trabecular metal tibial monoblock component.

机构信息

School of Biomedical Engineering, Dalhousie University, Halifax, Canada.

出版信息

J Arthroplasty. 2010 Sep;25(6):893-900. doi: 10.1016/j.arth.2009.06.015. Epub 2009 Aug 18.

Abstract

Radiostereometric analysis is a highly accurate technique that can be used for measuring micromotion at the bone-implant interface. The purpose of this study was to compare the inducible displacement of the uncemented Trabecular Metal (TM; Zimmer, Warsaw, Ind) tibial monoblock component with that of a cemented implant. Inducible displacement of 14 uncemented TM components and 11 cemented components was measured 24 to 48 months postoperatively. Longitudinal migration of the implants was also measured with radiostereometric analysis at 6, 12, and 24 months postoperatively. The uncemented TM group had significantly lower inducible displacement than the cemented components. Significant correlations were found between longitudinal migration and the inducible displacement tests. The low values of inducible displacement in the TM group indicated good fixation and a promising long-term prognosis.

摘要

放射学分析是一种非常精确的技术,可用于测量骨-植入物界面的微动。本研究的目的是比较非骨水泥 Trabecular Metal(TM;Zimmer,Warsaw,Ind)胫骨整体式组件与骨水泥固定植入物的诱导位移。术后 24 至 48 个月测量了 14 个非骨水泥 TM 组件和 11 个骨水泥组件的诱导位移。术后 6、12 和 24 个月还使用放射学分析测量了植入物的纵向迁移。非骨水泥 TM 组的诱导位移明显低于骨水泥组件。在纵向迁移和诱导位移测试之间发现了显著的相关性。TM 组的低诱导位移值表明固定良好,具有良好的长期预后。