An outbreak of Acinetobacter respiratory tract infection resulting from incomplete disinfection of ventilatory equipment.

作者信息

Cefai C, Richards J, Gould F K, McPeake P

机构信息

Department of Microbiology, Freeman Hospital, Newcastle upon Tyne.

出版信息

J Hosp Infect. 1990 Feb;15(2):177-82. doi: 10.1016/0195-6701(90)90128-b.

Abstract

An outbreak of multiresistant Acinetobacter anitratus respiratory tract infection occurred over a 30-day-period in six patients admitted to an intensive care unit. Standard precautions failed to curtail the outbreak, which was finally attributed to ineffective disinfection of reusable ventilator tubing.