Wrońska-Sewruk Agnieszka, Nestorowicz Andrzej, Kowalczyk Michał
I Klinika Anestezjologii i Intensywnej Terapii UM w Lublinie.
Anestezjol Intens Ter. 2009 Apr-Jun;41(2):73-7.
In recent years, different supraglottic airway devices became popular and new constructions have been proposed. We compared a classic laryngeal mask airway (LMA-classic) with the COBRA-PLA device (a LMA of different design).
Fifty adult ASA 1 and 2 adult patients, scheduled for minor urological interventions were randomly allocated to receive the LMA-Classic or the COBRA-PLA.
Time to secure airway was shorter for the LMA-Classic (16.8+/- 5 sec vs 33.0 +/- 19.6 sec; p<0.0001). The leak pressure was higher in the COBRA group (29.0+/- 7.5 vs 22.2 +/- 3.5 cm H2O; p=0.001). The cuff pressure, necessary for obtaining adequate seal, was higher for the LMA (83.6 +/- 14.1 vs 60.2 +/- 16.4 cm H2O; p<0.0001). We did not observe statistically significant differences in heart rate, arterial blood pressure, pulse oximetry and frequency of sore throat.
The LMA-classic was easier to insert but the COBRA device had more effective seal. These differences were not clinically important; both devices were found equally effective.
近年来,不同的声门上气道装置开始流行,并出现了新的构造。我们将经典喉罩气道(LMA-经典型)与COBRA-PLA装置(一种不同设计的喉罩)进行了比较。
50例计划进行小型泌尿外科手术的成年ASA 1级和2级患者被随机分配接受LMA-经典型或COBRA-PLA。
LMA-经典型建立气道的时间更短(16.8±5秒对33.0±19.6秒;p<0.0001)。COBRA组的漏气压力更高(29.0±7.5对22.2±3.5厘米水柱;p=0.001)。获得足够密封所需的套囊压力,LMA更高(83.6±14.1对60.2±16.4厘米水柱;p<0.0001)。我们未观察到心率、动脉血压、脉搏血氧饱和度和咽痛发生率有统计学显著差异。
LMA-经典型更容易插入,但COBRA装置的密封效果更好。这些差异在临床上并不重要;两种装置同样有效。