Department of Anesthesiology, University Hospital Lausanne, Lausanne, Switzerland.
J Cardiothorac Vasc Anesth. 2010 Apr;24(2):250-6. doi: 10.1053/j.jvca.2009.05.029. Epub 2009 Aug 22.
The measurement of cardiac output is a key element in the assessment of cardiac function. Recently, a pulse contour analysis-based device without need for calibration became available (FloTrac/Vigileo, Edwards Lifescience, Irvine, CA). This study was conducted to determine if there is an impact of the arterial catheter site and to investigate the accuracy of this system when compared with the pulmonary artery catheter using the bolus thermodilution technique (PAC).
Prospective study.
The operating room of 1 university hospital.
Twenty patients undergoing cardiac surgery.
CO was determined in parallel by the use of the Flotrac/Vigileo systems in the radial and femoral position (CO_rad and CO_fem) and by PAC as the reference method. Data triplets were recorded at defined time points. The primary endpoint was the comparison of CO_rad and CO_fem, and the secondary endpoint was the comparison with the PAC.
Seventy-eight simultaneous data recordings were obtained. The Bland-Altman analysis for CO_fem and CO_rad showed a bias of 0.46 L/min, precision was 0.85 L/min, and the percentage error was 34%. The Bland-Altman analysis for CO_rad and PAC showed a bias of -0.35 L/min, the precision was 1.88 L/min, and the percentage error was 76%. The Bland-Altman analysis for CO_fem and PAC showed a bias of 0.11 L/min, the precision was 1.8 L/min, and the percentage error was 69%.
The FloTrac/Vigileo system was shown to not produce exactly the same CO data when used in radial and femoral arteries, even though the percentage error was close to the clinically acceptable range. Thus, the impact of the introduction site of the arterial catheter is not negligible. The agreement with thermodilution was low.
心输出量的测量是评估心功能的关键要素。最近,一种无需校准的基于脉搏轮廓分析的设备(FloTrac/Vigileo,Edwards Lifescience,加利福尼亚州欧文)问世。本研究旨在确定动脉导管位置是否会对其产生影响,并研究该系统与使用热稀释法的肺动脉导管(PAC)相比的准确性。
前瞻性研究。
1 所大学医院的手术室。
20 名接受心脏手术的患者。
使用 Flotrac/Vigileo 系统分别在桡动脉和股动脉位置(CO_rad 和 CO_fem)和 PAC 同时测量心输出量,并将其作为参考方法。在定义的时间点记录数据三联体。主要终点是比较 CO_rad 和 CO_fem,次要终点是与 PAC 的比较。
共获得 78 个同步数据记录。CO_fem 和 CO_rad 的 Bland-Altman 分析显示,偏差为 0.46 L/min,精度为 0.85 L/min,误差百分比为 34%。CO_rad 和 PAC 的 Bland-Altman 分析显示,偏差为-0.35 L/min,精度为 1.88 L/min,误差百分比为 76%。CO_fem 和 PAC 的 Bland-Altman 分析显示,偏差为 0.11 L/min,精度为 1.8 L/min,误差百分比为 69%。
即使误差百分比接近临床可接受范围,FloTrac/Vigileo 系统在桡动脉和股动脉中使用时也不会产生完全相同的心输出量数据。因此,动脉导管插入部位的影响不可忽视。与热稀释法的一致性较低。