Suppr超能文献

在使用Look-Locker技术的三维dGEMRIC中,通过局部翻转角校正改善容积T1定量

Local flip angle correction for improved volume T1-quantification in three-dimensional dGEMRIC using the Look-Locker technique.

作者信息

Siversson Carl, Tiderius Carl-Johan, Dahlberg Leif, Svensson Jonas

机构信息

Department of Radiation Physics, Lund University, Malmö, Sweden.

出版信息

J Magn Reson Imaging. 2009 Oct;30(4):834-41. doi: 10.1002/jmri.21906.

Abstract

PURPOSE

To present an evaluation method for three-dimensional Look-Locker (3D-LL) based T1 quantification, calculating correct T1 values independent of local flip angle (FA) variations. The method was evaluated both in phantoms and in vivo in a delayed Gadolinium Enhanced MRI of Cartilage (dGEMRIC) study with 33 subjects.

MATERIALS AND METHODS

T1 was measured with 3D-LL, using both local FA correction and a precalculated FA slice profile, and compared with standard constant FA correction, for all slices in phantoms and in both femur condyles in vivo. T1 measured using two-dimensional Inversion Recovery (2D-IR) was used as gold standard.

RESULTS

Due to the FA being slice dependent, the standard constant FA correction results in erroneous T1 (systematic error = 109.1 ms in vivo), especially in the outer slices. With local FA correction, the calculated T1 is excellent for all slices in phantoms (<5% deviation from 2D-IR). In vivo the performance is lower (systematic error = -57.5 ms), probably due to imperfect inversion. With precalculated FA correction the performance is very good also in vivo (systematic error = 13.3 ms).

CONCLUSION

With the precalculated FA correction method, the 3D-LL sequence is robust enough for in vivo dGEMRIC, even outside the centermost slices.

摘要

目的

提出一种基于三维Look-Locker(3D-LL)的T1定量评估方法,该方法能够在不考虑局部翻转角(FA)变化的情况下计算出正确的T1值。在33名受试者的软骨延迟钆增强磁共振成像(dGEMRIC)研究中,对该方法在体模和体内进行了评估。

材料与方法

使用3D-LL测量T1,采用局部FA校正和预先计算的FA切片轮廓,并与标准恒定FA校正进行比较,对体模中的所有切片以及体内双侧股骨髁进行比较。使用二维反转恢复(2D-IR)测量的T1用作金标准。

结果

由于FA依赖于切片,标准恒定FA校正会导致T1错误(体内系统误差 = 109.1 ms),尤其是在外围切片中。采用局部FA校正时,体模中所有切片计算出的T1都非常出色(与2D-IR的偏差<5%)。在体内性能较低(系统误差 = -57.5 ms),可能是由于反转不完善。采用预先计算的FA校正时,在体内性能也非常好(系统误差 = 13.3 ms)。

结论

采用预先计算的FA校正方法,3D-LL序列对于体内dGEMRIC足够稳健,即使在最中心切片之外也是如此。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验